1. Instrument Safety
1.1. Correct Use
The ALTAIR 5 and ALTAIR 5 IR Multigas Detectors are for use by
trained and qualified personnel. They are designed to be used when
performing a hazard assessment to:
•
Assess potential worker exposure to combustible and toxic gases
and vapors as well as low level of oxygen.
•
Determine the appropriate gas and vapor monitoring needed for a
workplace.
The ALTAIR 5 and ALTAIR 5 IR Multigas Detectors can be equipped
to detect up to five gases in the ALTAIR 5 and up to six gases in the
ALTAIR 5 IR:
•
Combustible gases and certain combustible vapors
•
Oxygen-deficient or oxygen-rich atmospheres
•
Specific toxic gases for which a sensor is installed.
While the instrument can detect up to 25 % oxygen in ambient air, it is
approved for use only up to 21 % oxygen.
The ALTAIR 5 IR Gas Detector contains one Infrared sensor to detect
either CO 2 or specific combustible gases up to 100% volume, in
addition to up to five other gases.
Read and follow all instructions carefully.
•
Perform a blocked flow test before each day's use if
equipped with an integral pump.
•
Check calibration before each day's use and adjust if
necessary.
•
Check calibration more frequently if exposed to silicone,
silicates, lead-containing compounds, hydrogen sulfide, or
high contaminant levels.
•
Recheck calibration if unit is subjected to physical shock.
•
Use only to detect gases/vapors for which a sensor is
installed.
•
Do not use to detect combustible dusts or mists.
" WARNING
1-1