Resetting Default Parameters; Settings Menu - aqua DGTEC CD Mode D'emploi

Masquer les pouces Voir aussi pour DGTEC CD:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 70
dGdoser Model
Programmable digital peristaltic pump
3.3.6 R
D
ESETTING
EFAULT
From the r.P.d entry you will be able to reset the equipment default parameters (refer to the Default Parameters
Table in Attachment F). If you choose to reset the initial values, you will go back to the initial display. See Fig. 17.
Once you confirm with YES, you cannot go back.
3.3.7 S
M
ETTINGS
ENU
From the InP entry of the Programming Menu, via the Enter key, you can access the Settings Menu.
From the Settings Menu you can carry out the following operations (refer to Figure 18):
 Set the password to access the Programming section, PAS: with Enter you can go to the "000" selection
page, with the ▲ and ▼ keys you can increase/decrease the value displayed and with Enter you can scroll
through the figure you intend to set. After choosing the new password, press Enter for 3 seconds to go back
to the PAS entry;
 Enable the pump stop in the case of a St.P level alarm: to access and exit the submenu, use Enter, to scroll
through "YES/no" use the ▲ and ▼ keys;
 Choose whether to disable the over-dosage alarm OFA OF.r only from the switch by placing it in the OFF
position, or to disable it from the switch as well as when turning off the equipment, Aut: to enter and exit the
sub-menu, use the Enter key, to scroll "OFF/Aut" use ▲ and ▼;
 Select the time frame to control the OFA alarm, OFA, within a range of 1-999 seconds or disable it (OFF): to
enter and exit the submenu use Enter, to scroll through the values use the ▲ and ▼ keys;
 Set the time of the pump maintenance alarm, O.F.d, from 0 (disabled) to 999 hours, based on the following
format:
- h.mm up to 9 h and 59 m
- hh.m up to 99 h and 59 m (minutes expressed in tens)
- hhh up to 999 h
To enter and exit the submenu use Enter and increase/decrease the value with the ▲ and ▼ keys;
 Choose if you want to make the Set-point, OnL, a value that can be changed on-line or from the User Menu:
to enter and exit the sub-menu use Enter, to scroll through "YES/no" use the ▲ and ▼ keys;
 Enable the A.bu buzzer when an alarm is activated: to enter and exit the sub-menu use Enter, to scroll
through "YES/no" use the ▲ and ▼ keys;
 Set the temperature coefficient value C.d.t (see Operation, Section 3.1) from 0.0 to 5.0 (percentage value
from 0.0% to 5.0%): to enter and exit the sub-menu use the Enter key, the value is increased/decreased by
using the ▲ and ▼ keys;
 Set the opening time of the solenoid valve t.on, in the case of dosage without the peristaltic pump (solid
detergent), in the range of 0-30 seconds: to enter and exit the sub-menu, use the Enter key, the value is
increased/decreased by using the ▲ and ▼ keys;
ADSP7000634
P
ARAMETERS
Fig. 17- Default parameters reset
rev. 1.2
INSTRUCTION AND MAINTENANCE MANUAL
30/06/2014
ENGLISH
51/206

Publicité

Table des Matières
loading

Table des Matières