Basic Operating Modes; Milliampere Source - WIKA CEP1000 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6. Commissioning, operation

6.2 Basic operating modes

The CEP1000 has 4 operating modes for the mA range and one operating mode for
voltage measurement. The following text and figures explain different operating modes.

6.2.1 Milliampere source

Figure 3 shows a typical application, for which the CEP1000 is used as a source of
current for the range from 4 ... 20 mA for the instrument to be tested.
CEP 1000
SOURCE
MODE
%ERROR
LOOPPWR
Calibration Line
1) Turn on the instrument. As a standard the operating mode source of current in the
mA range is set.
2) Connect the instrument to be calibrated.
For the instrument to function as a source of current, the circuit must be
closed. In case of an interrupted circuit the LCD display "OL" is blinking
and indicating an overload or an interruption of the circuit.
WIKA operating instructions, model CEP1000
4 ... 20 mA
mA
Max. load:
MAX
AUTO
MIN
25%
Figure 3 - Current source for 4 ... 20 mA
( + )
Instrument with an input
of 4 ... 20 mA
( - )
1200 Ω without HART™
950 Ω with HART™
GB
15

Publicité

Table des Matières
loading

Table des Matières