Accuracy
Environmental
Ambient
Operating
Ranges
Storage
Temperature
General
Low battery
Indication
Measurement
rate
Accuracy
Battery
Battery Life
Dimensions
Weight
Accuracy is specified for operating
temperatures over the range of 18°C
to 28°C (64°F to 82°F), for 1 year, not
including thermocouple error.
±(0.1%rdg+1°C) on -60°C to 1372°C
±(0.1%rdg+2°C) on -60°C to -200°C
±(0.1%rdg+2°F) on -76°F to 1999°F
±(0.1%rdg+4°F) on -76°F to -328°F
0°C to 50°C (32°F to 122°F) <80% R.H.
-20°C to 60°C (-4°F to 140°F) <80% R.H.
with battery removed from meter.
The "
" is displayed when the battery
voltage drops below the operating level.
2.5 times/second.
Stated accuracy at 23°C±5°C, <75% R.H.
1.5V x 4pcs AAA size
200 hours typical with carbon zinc
battery.
160mm(H) x 83mm(W) x 38mm(D);
6.3 in (H) x 3.3 in (W) x 1.5 in (D).
Approx. 240g (0.5lb) including batteries.
6