4. Install a replacement fuse, and reassemble the tip.Ensure the retainer ring is properly
seated and tightened (contact your equipment provider for replacement part).
Cigarette Adapter Plug
CAUTION For continued protection against risk of fire, only use fuse specified.
OxyGo Column Change Procedure
NOTE
Column change instructions are only to be used when maintenance is
required and are not intended for practice purpose.
1. Turn off the OxyGo concentrator by pressing the power button to shut down the
device.
2. Remove the OxyGo concentrator from the carry bag.
3. Remove the battery from the OxyGo concentrator.
4. Turn OxyGo over so that it is standing in the upside
down position where the display panel is
facing down and the bottom of the OxyGo is
facing up.
5. There are two columns (metal tubes) on the
OxyGo concentrator and one is located on
each side of the device.
Fuse
Retainer Tip
Cleaning, Care and Maintenance Chapter 6
Retainer
33