3
General Instructions
1. Place the OxyGo in a well ventilated location.
Air intake and exhaust must have clear access.
Locate the OxyGo in such a way that any
auditory alarms may be heard.
WARNING Avoid use of the OxyGo in presence of pollutants, smoke or fumes. Do not
use the OxyGo in presence of flammable anesthetics, cleaning agents or
C
!
other chemical vapors.
CAUTION
Do not obstruct air intake or exhaust when operating the device.
Blockage of air circulation or proximity to a heat source may lead to
internal heat buildup and shutdown or damage to the concentrator.
CAUTION
The OxyGo Concentrator is designed for continuous use. For optimal sieve
bed (columns) life, the product should be used frequently.
2. Ensure both particle filters are in place.
CAUTION
Do not operate the OxyGo without both particle
filters in place. Particles drawn into the system may
damage the equipment.
Operating Instructions
Intake
Exhaust
Operating Instructions Chapter 3
Exhaust
Intake
11