Filter Cleaning and Replacement
Both particle filters must be cleaned weekly to
ensure the ease of air flow. Remove filters from
the front of the device. Clean the particle filters
with a mild liquid detergent (such as Dawn
and water; rinse in water and dry before reuse.
NOTE
To purchase additional particle filters contact your equipment provider.
Output Filter
The output filter is intended to protect the user from inhalation of small particles in
the product gas flow. The OxyGo includes an output filter conveniently located behind
the removable cannula nozzle fitting. Manufacturer requires that this filter be replaced
between patients.
The output filter may be replaced by the equipment provider or by the owner using the
Output Filter Replacement Kit (1400-1322).
The OxyGo Concentrator must be cleaned and disinfected as per the above instructions
for each new patient. No special maintenance needs to be carried out by the patient.
Your equipment provider performs maintenance operations to assure continued
reliable service from your OxyGo. The manufacturer's instructions for the preventative
maintenance of the devices are defined in the service manual. All work must be
performed by trained technicians certified by the manufacturer.
DC Input Cable Fuse Replacement
The Cigarette Lighter DC power plug contains a fuse. If the DC input cable is being used
with a known good power source and the unit is not receiving power, the fuse may need
to be replaced.
To replace the fuse, follow these instructions and refer to the photograph below.
1. Remove the tip by unscrewing the retainer. Use a tool if necessary.
2. Remove the retainer, tip and fuse.
3. The spring should remain inside the Cigarette Lighter Adapter housing. If the spring
is removed, make sure to replace the spring first before inserting the replacement
32
fuse.
It may be necessary to clean the particle filters more often in dusty
environments.
)