10.
Trouble Shooting
FAULT SYMPTOM
"No display"
"Display unstable"
"No signal output"
"Display too low"
MESSAGE ON DISPLAY
"Configuration error"
"HW-probe error"
"Probe error"
"Temperature sensor error"
"TKα and δT error"
With an error in the conductivity measuring circle at the current outputs for conductivity
NOTE
there are 0 mA. (only at 4...20 mA!).
If a fault cannot be eliminated with the list above, please send the transmitter with the
measuring cell to the factory.
NOTE
After a repair in the factory the unit will be set to the ground configuration (chapter 8.4.5).
417101109_LMIT08.doc
CAUSE/MALFUNCTIOM
Display does not light up, no supply
voltage fuse is working
Air bubbles in installation site of
measuring cell Insufficient pipe
filling
Supply voltage unstable
Wire/terminals defective or
connected analysing unit faulty
Measuring hole contaminated
Storage error
Short circuit in the transducer
Transducer defect or wrong
connected
Temperature sensor defect or
wrong connected
Limits of the temperature
compensation exceeded
- 79 -
RECTIFICATION
Check supply voltage or check
whether miniature or fuse faulty if nec.
replace fuse
Check measuring cell installation site
for air bubbles or inadequate pipe
filling
Check supply voltage for fluctuations
Check wire, terminals and connected
analysing unit
Check zero point suppression
(conductivity < than start of
measurement)
Clean measuring hole
Push 2 x SEL button; if fault comes
once more exchange the unit
Exchange transducer
Check connection (core wire colour
code) possibly exchange transducer
Reduce TKα or temperature to values
within the limits
Rev. 3-01.10