Bedienungsanleitung Operating Instructions Notice technique LMIT 08 Transmitter für elektrodenlose Leitfähigkeitsmessung und Temperaturmessung Transmitter for electrodeless conductivity measurement and temperature measurement Transmetteur pour le mesurage sans électrode de la conductivité et de la température 417101109_LMIT08.doc - 1 - Rev. 3-01.10...
Page 4
TABLE OF CONTENS General ..............................50 Scope of guarantee ..............................50 Contact address / Manufacturer ..........................50 Safety instructions..........................51 Pointers ..................................51 Numberings ................................51 Delivery scope............................52 Installation instructions........................53 Structure ............................... 54 Clamping ring version .............................54 Flange version ................................55 Functional Discription ......................... 56 Equipment versions ..............................56 6.1.1 BASIC Version................................56...
Page 5
Calibration functions ...............................72 8.5.1 Functional choice in calibration mode........................72 8.5.2 Setting the assembly factor ............................73 8.5.2.1 Pharma version ....................................73 8.5.3 Automatic TCα determination ..........................73 8.5.4 Calibration of temperature sensor ..........................74 8.5.5 Calibrate measuring cell with reference resistance ....................74 Display selection in "Set CIP ranges" mode ......................76 8.6.1 Ranges selection menu ............................76 8.6.2...
Page 6
TABLE DES MATIÈRES Informations générales........................92 L’étendue de notre garantie :..........................92 Adresse de contact / Fabricant ..........................92 Instructions de sécurité........................93 Avertissements ...............................93 Enumérations .................................93 Contenu de la livraison........................94 Indications concernant l’installation....................95 Composants ............................96 Version à clamp ..............................96 Flanschversion ...............................97 Description fonctionnelle ........................
Page 7
8.4.5 Réduire configuration............................113 8.4.5.1 Configuration usine ..................................114 Fonctions de calibrage............................114 8.5.1 Choix de fonction dans le mode de calibrage .......................114 8.5.2 Régler facteur d’installation ..........................115 8.5.2.1 Version Pharma.................................... 115 8.5.3 Détermination TKα automatique...........................115 8.5.4 Calibrage de la sonde de température........................116 8.5.5 Calibrer le capteur avec résistance de référence ....................116 Sélection d’affichages dans le mode «...
Bei Reparaturen werden nur Original-Ersatzteile verwendet. Im Übrigen gelten die allgemeinen Garantie- und Leistungsbedingungen der ECOLAB-Engineering GmbH. Kontaktadresse / Hersteller ECOLAB-Engineering GmbH Raiffeisenstraße 7 D-83313 Siegsdorf Telefon (+49) 86 62 / 61 0 Telefax (+49) 86 62 / 61 2 35 eMail: engineering-mailbox@ecolab.com...
Sicherheitshinweise Der Leitfähigkeits- und Temperaturtransmitter LMIT 08 darf nur mit 24 V Gleich- oder Wechselspannung betrieben werden. Reparaturarbeiten dürfen nur am abgeschalteten und druckfreien Gerät durchgeführt werden. Dieses Messgerät ist gemäß EN 61010-1, Teil 1 gebaut und geprüft, und hat das Werk in sicherheitstechnisch einwandfreiem Zustand verlassen.
Abb. 4.1 Die Strömung muss von unten nach oben erfolgen (siehe Abb. 4.1). Hierbei empfiehlt es sich, das Gehäuse des LMIT 08 so zu drehen, dass die elektrischen Anschlüsse nach unten zeigen. Abb. 4.2 Sollte der Transmitter in waagrecht verlaufenden Rohrleitungen installiert werden, muss der Einbau von unten erfolgen (siehe Abb.
Aufbau Spannringversion Abb. 5.1 Pos. Bezeichnung Hilfsenergie 24 V AC/DC Stromausgänge 0(4)…20 mA LF + Temp. Transmitter mit integriertem Messwertgeber (Kompaktversion) zum Einbau in Rohrleitungen. Abb. 5.2 Pos. Bezeichnung Hilfsenergie 24 V AC/DC Stromausgänge 0(4)…20 mA LF + Temp. Potentialausgleich min. 1,5 mm² Transmitter mit separatem Messwertgeber (Wandaufbau) zum Einbau in Rohrleitungen.
Flanschversion Abb. 5.5 Pos. Bezeichnung Hilfsenergie 24 V AC/DC Stromausgänge 0(4)…20 mA LF + Temp. Transmitter mit integriertem Messwertgeber (Kompaktversion) zum Einbau in Rohrleitungen. Abb. 5.6 Pos. Bezeichnung Hilfsenergie 24 V AC/DC Stromausgänge 0(4)…20 mA LF + Temp. Potentialausgleich min. 1,5 mm² Transmitter mit separatem Messwertgeber (Wandaufbau) zum Einbau in Rohrleitungen.
6.1.2 CIP-Ausführung Der Leitfähigkeits-/Temperaturtransmitter LMIT 08 beinhaltet in der Ausführung "CIP" alle für die Phasentrennung von leitfähigen Medien notwendigen Funktionen. Geräteausführungen: Kompaktgerät mit integriertem Messwertgeber zum Rohreinbau Wandaufbaugerät mit separatem Messwertgeber zum Rohreinbau Die CIP-Ausführung enthält alle Möglichkeiten der BASIS-Ausführung, zusätzlich dazu: •...
6.1.4 Pharma-Ausführung Der Leitfähigkeits-/Temperaturtransmitter LMIT 08 beinhaltet in der Ausführung "Pharma" alle Funktionen der Basis- und CIP-Ausführung. Geräteausführungen: Kompaktgerät mit integriertem Messwertgeber zum Rohreinbau Wandaufbaugerät mit separatem Messwertgeber zum Rohreinbau Besonderheit: Ein optimiertes Sensorsystem steigert die Messgenauigkeit im untersten µS-Messbereich gegenüber dem Standardgerät.
Abb. 7.7 Typ: Anschweißflansch zum Tankwandeinbau Material: 1.4404, Dichtung EPDM Abb. 7.8 Typ: Durchflußarmatur Material: 1.4404,Dichtung EPDM 7.1.1 Transmitter mit integriertem Messwertgeber (Kompaktversion) Die Querbohrung der Messkalotte ist grundsätzlich parallel zur Rohrachse, d.h. in Fließrichtung des Mediums auszurichten. ACHTUNG Abweichungen können zu Verfälschungen der Messwerte führen (Wirbelbildung). Eine Kontrolle der Einbaulage ist anhand von Markierungskerben auf der Messkalotte möglich.
7.1.2 Transmitter mit separatem Messwertgeber (Wandaufbauversion) Der Transmitter ist für Wandbefestigung geeignet. Die Befestigung erfolgt mit zwei Schrauben. Die Querbohrung der Messkalotte ist grundsätzlich parallel zur Rohrachse, d.h. in Fließrichtung des Mediums auszurichten. Abweichungen können zu Verfälschungen der Messwerte führen (Wirbelbildung). Eine ACHTUNG Kontrolle der Einbaulage ist anhand von Markierungskerben auf der Messkalotte möglich.
2) Leitungen zueinander verdrillt (schwarz/rot) 7.2.1 Hilfsenergie Anschluss Nach EN 60204 Abschnitt 8 ist am LMIT 08 neben den Anschlüssen für die Versorgungsspannung X1/1, 2 auch ein Anschluss für den Schutzleiter vorhanden. Beim Transmitter in Wandaufbau-Ausführung kann der am Befestigungsflansch vorhandene Anschluss für den Potentialausgleich auch zum Anschluss des Schutzleiters verwendet...
7.2.6 Klemmenbelegungsplan PROFIBUS-Modul DPV1 und PA Abb. 7.14 Pos. Bezeichnung PROFIBUS Anschluss Der PROFIBUS-Anschluss für das LMIT 08 wird von einem T-Stück der PROFIBUS-Leitung abgezweigt. Die max. Leitungslänge der Stichleitung beträgt 1 m. Der Schirm der Busanschlussleitung muss im LMIT 08 isoliert geführt werden.
Anschluss Abb. 8.1 Pos. Bezeichnung LC Display Kontrasteinstellung für Display Programmspeicher Übergangsstecker für CIP- bzw. PROFIBUS-Modul Bedientasten Pos. Symbol Funktion NEXT Anwahl der Bedienungsebenen Angezeigten Wert nach oben verändern Anwahl eines Untermenüs Angezeigten Wert nach unten verändern Anwahl eines Untermenüs Bestätigung einer Eingabe SEL/ Abspeichern des angezeigten...
8.2.3 Anzeige der unkompensierten Leitfähigkeit Abb. 8.4 Mit der SEL-Taste kann die unkompensierte d.h. die tatsächlich gemessene Leitfähigkeit zur Anzeige gebracht werden. Der Stromausgangswert für die Leitfähigkeit ändert sich ebenfalls bezogen auf den HINWEIS unkompensierten LF-Wert. 8.2.4 TKα- und Referenztemperatur-Anzeige (nur in den Leitfähigkeits-Messbereichen) Abb.
8.4.1 Anzeigen in der Konfigurationsebene Abb. 8.9 Die Menüpunkte werden mit den Pfeiltasten angewählt und mit der SEL-Taste aktiviert. 8.4.2 Konfiguration abfragen 8.4.2.1 Messbereichsabfrage Abb. 8.10 8.4.2.2 Stromausgangssignal-zuordnung für Leitfähigkeit bzw. Konzentration Abb. 8.11 8.4.2.3 Temperaturkoeffizient (nur in den LF-Mess-bereichen) Abb.
8.4.2.6 Geräte-Nr.-Abfrage Abb. 8.15 8.4.3 Konfiguration ändern 8.4.3.1 Codezahlschutz Abb. 8.16 Der Zugriff auf die Konfiguration des Gerätes wird bei richtiger Eingabe der Codezahl freigegeben. Zur Inaktivierung des Codezahlschutzes muss ebenfalls die richtige Codezahl eingegeben werden. Ab Werk ist der Codezahlschutz nicht aktiviert. Bei erstmaliger Aktivierung des Codezahlschutzes kann die Codezahl zwischen 001 und 999 frei gewählt werden.
8.4.3.2 Messbereich ändern Abb. 8.18 Wird die Funktion "Messbereich ändern" aufgerufen, erscheint das abgebildete Menü. Abhängig davon, ob vorher ein Leitfähigkeits- oder Produkt-Messbereich eingestellt war, wird dieser in der zweiten Zeile angezeigt. Mit den Pfeiltasten wird zwischen Konzentrations- und LF-Messung gewählt. Mit der SEL-Taste wird eine der nachfolgenden Funktionen aufgerufen. 8.4.3.2.1 Ändern des Leitfähigkeits-Messbereiches Abb.
8.4.3.5 Zuordnung des LF-Stromausgangssignales ändern (Nullpunktunterdrückung und Messbereichsspreizung) Der Transmitter LMIT 08 wurde so konzipiert, dass in der Standardeinstellung dem Stromausgangssignal 0(4)...20 mA der Messbereichsanfangs- und endwert zugeordnet sind. Darüber hinaus ist die Möglichkeit gegeben, eine optimale Bereichsspreizung und Nullpunktunterdrückung entsprechend der Messaufgabe durchzuführen. Dies kann nötig sein, wenn beispielsweise geringe Messsignaländerungen mit einer großen Auflösung erfasst...
Einstellgrenzen: - Nullpunkt: 0...50 °C - SPAN: 50...150 °C 8.4.3.7 Gerätenummer ändern Abb. 8.28 Jedem LMIT 08 kann eine dreistellige individuelle Gerätenummer zugeordnet werden. Einstellbereich: 1...999, bzw. --- falls keine Gerätenummer zugeordnet werden soll. 417101109_LMIT08.doc - 28 - Rev. 3-01.10...
8.4.4 Konfigurationswerte ausdrucken Abb. 8.29 Die Druckerschnittstelle RS232 erlaubt den Ausdruck von Konfigurations- und Parametrierdaten. Eine Online-Protokollierung von Leitfähigkeits- und Temperaturwerten ist nicht möglich. Falls der Ausdruck nicht korrekt erfolgt, Druckeranschluss und Druckerstatus (online) überprüfen. Schnittstellenparameter: Baudrate 9600 Parität keine Datenbits Stopbit Protokoll...
Der Gefäßfaktor ist ein Maß für die Beeinflussung der Zellenkonstante der Messzelle durch die Einbausituation. Der Gefäßfaktor erlaubt eine Anpassung der LMIT 08-Kalibrierung an die verschiedenen Einbausituationen. Die untenstehende Tabelle zeigt die notwendigen Gefäßfaktoren für Tankeinbau- und Durchflussarmaturen, die mit dem LMIT 08 ausgeliefert werden. Abb. 8.33 Armatur Gefäßfaktor Tankeinbau 1,000...
Abb. 8.38 Bei Auftreten eines Fehlers: Abb. 8.39 8.5.5 Messzelle kalibrieren mit Referenzwiderstand Der Leitfähigkeits-/Temperaturtransmitter LMIT 08 ist ab Werk mit der mitgelieferten Messzelle kalibriert. Sollte nach einem Austausch des Messwertgebers eine Neukalibrierung nötig sein, wird wie folgt vorgegangen: •...
Page 33
Kalibriervorganges wird für 1 sec. die Meldung "Kalibrierung OK" ausgegeben und dann auf den nächsten Kalibrierwert weitergeschaltet. Abb. 8.42 Um die Messgenauigkeit des LMIT 08 zu gewährleisten, müssen die Schleifenwiderstände HINWEIS eine Genauigkeit von mindestens 0,2 % aufweisen. Dies ist bei dem als Zubehör erhältlichen LF-Simulator (289190 und 289191) gewährleistet.
Das Messgerät ist gegen Störeinflüsse entsprechend den NAMUR-Empfehlungen NE21 geschützt. Wartungsarbeiten Die Wartungsarbeit am LMIT 08 besteht im Wesentlichen darin, in gewissen Abständen die Querbohrung der Messkalotte auf Ablagerungen zu überprüfen, und falls erforderlich zu reinigen. Die elektrische Prüfung des...
Page 37
Störungsüberprüfung FEHLER-SYMPTHOM URSACHE/STÖRUNG BEHEBUNG Prüfen, ob die Display leuchtet nicht, keine Versorgungsspannung vorhanden „Keine Anzeige“ Versorgungsspannung bzw. ist, bzw. prüfen ob die Sicherung defekt Feinsicherung in Ordnung ist, ggf. Sicherung austauschen Luftblasen am Einbauort der Einbauort der Messzelle auf Messzelle ungenügende Luftblasen bzw.
Page 38
Dichteinsatz für Kabel ø 6,5 - 4,0 "gelb" 417804027 CIP-Platine 289152 Platine PROFIBUS-DPV1 289156 Platine PROFIBUS-PA 289158 Fenster bedruckt LMIT 08 38910101 Standardmesszellen können nicht mit Pharmageräten und Pharmamesszellen können ACHTUNG nicht mit Standardgeräten kombiniert werden. 417101109_LMIT08.doc - 38 - Rev. 3-01.10...
415501232 Spannringsatz, 1.4404 415501231 Durchflussgehäuse mit Flanschbefestigung, 1.4404 Gehäuse mit 2 Anschlusssystemen eine Seite für Anschluss des LMIT 08 Messwertgebers, andere Seite mit lösbarem Blindflansch DN 40 da 41 x 1,5 (AD x S) 415501261 DN 50 da 53 x 1,5 (AD x S)
Stromversorgung Versorgungsspannung: 24 V AC oder DC Toleranz: +/- 15 % Leistungsaufnahme: ca. 6 VA Absicherung: 5 x 20 mm Feinsicherung 400 mA, träge Elektrischer Anschluss: steckbare Schraubklemmen Umgebungsbedingungen Zulässige Umgebungstemperatur: 0 ... 50°C Einfluss Umgebungstemperatur: < 0,2 % / 10 K innerhalb des zulässigen Bereiches Einfluss Versorgungsspannung: <...
14.2 Diagramme 14.2.1 Spezifische Leitfähigkeit Spez. Leitfähigkeit æ verschiedener Elektrolytlösungen in Abhängigkeit von der Konzentration bei 20 °C 1000 NaOH NaOH NaCl NaCl 0.6 0.8 1 Konzentration in [Gew. %] 14.2.2 Temperaturkoeffizienten Zusammenhang zwischen den Temperaturkoeffizenten bei verschiedenen Elektrolytlösungen (Referenztemperatur 20 °C) NaCl NaCl NaOH...
Zusammenhang zwischen den Temperaturkoeffizienten bei den Referenztemperaturen 0 °C und 25 °C bezogen auf 20 °C. TK- α (0 TK- α (0 TK- α (25 TK- α (25 Temperaturkoeffizent α (20 °C) in [% / K] 14.2.3 Zulässiger Betriebsdruck der Messzelle Betriebstemperatur in [°C] HINWEIS Im Grenzbereich um 140 °C ist nur kurzzeitiger Betrieb zulässig.
14.3 Anschlussplan CIP 14.3.1 Beispiel für Signaleingang mit Relais 14.3.2 Beispiel für Signaleingang mit Optokoppler 417101109_LMIT08.doc - 45 - Rev. 3-01.10...
14.3.5 Anschlüsse im Verbund Um Querströme zwischen Leitfähigkeits- und Temperatur-Stromausgang zu vermeiden, müssen speicherprogrammierbare Steuerungen, Schreiber oder dergleichen, deren HINWEIS Minuseingänge untereinander verbunden sind, an letzter Stelle der Stromschleife eingefügt werden! Es dürfen in keinem Fall Strom-Schnittstellenkarten 4...20 mA zum Einsatz kommen, die eine ACHTUNG Spannung zur Versorgung von Zweidraht-Transmittern ausgeben! (Gefahr der Zerstörung der Stromausgänge am...
Page 48
14.5 Bedienschema abfragen: siehe 8.4.2 ändern: siehe 8.4.3 ausdrucken: siehe 8.4.4 zurücksetzen : siehe 8.4.5 zurück zur Messebene Gefäßfaktor: siehe 8.5.2 TKα automatisch: siehe 8.5.3 Temperaturfühler: siehe 8.5.4 mit Widerstand : siehe 8.5.5 zurück zur Messebene Adresse einstellen: siehe Bereich 1-4 siehe 8.6.2 zurück zur Messebene zurück zur Messebene...
The transducer is used in accordance with the instructions in the technical manual. • Only orginal spare parts are used for repairs. In addition, the general warranty and service conditions of the company ECOLAB-Engineering GmbH are applicable. Contact address / Manufacturer ECOLAB-Engineering GmbH Raiffeisenstraße 7...
Safety instructions The conductivity and temperature transmitter LMIT 08 may only run on a 24 V, DC or AC supply. Repairs must always be carried out on units when they are switched off and unpressurized. This transducer has been built and tested EN 61010-1, Part 1 and left the works in a safe and fully operational state.
Fig. 4.1 The flow has to be from below up (see Fig. 4.1). It is recommended to put the LMIT 08 the way that the electrical connections point down. Fig. 4.2...
Structure Clamping ring version Fig. 5.1 Pos. Description auxiliary power 24 V AC/DC current outputs 0(4)…20 mA cond. + Temp. Transmitter with integrated transducer (compact version) for pipe installation Fig. 5.2 Pos. Description auxiliary power 24 V AC/DC current outputs 0(4)…20 mA cond. + Temp. potential equalization min.
Flange version Fig. 5.5 Pos. Description auxiliary power 24 V AC/DC current outputs 0(4)…20 mA cond. + Temp. Transmitter with integrated transducer (compact version) for pipe installation Fig. 5.6 Pos. Description auxiliary power 24 V AC/DC current outputs 0(4)…20 mA cond. + Temp. potential equalization min.
Equipment versions 6.1.1 BASIC Version The BASIC version of the LMIT 08 conductivity / temperature transmitter contains all the necessary functions for monitoring and making up the concentration of conductive media. Equipment versions: Compact unit with integrated transducer for pipe installation...
6.1.2 CIP Version The conductivity / temperature transmitter LMIT 08 embodies all the functions necessary for separating the phases of conductive media. Equipment versions: Compact unit with integrated transducer for pipe installation Wall-mounted version with separate transducer for pipe installation The CIP version contains all the facilities of the BASIC version as well as the following: •...
6.1.4 Pharma Version The "Pharma" version of the LMIT 08 conductivity and temperature transmitter performs all functions that are available in the base and CIP versions. Versions available: Compact instrument with integrated transducer for installation into piping system Wall-mounted instrument with separate transducer for installation into piping...
Connection Mechanical connections Fig. 7.1 Compact unit with clamping ring Fig. 7.2 Wall mounted unit Fig. 7.3 Compact unit with flange Fig. 7.4 Type: Flow fitting (DIN 11850) Material: 1.4404, sealing EPDM Fig. 7.5 Type: weld-on-ring for tank wall mounting Material: 1.4404, sealing EPDM Abb.
Fig. 7.7 Type: weld-on-flange for tankwall mounting Material: 1.4404, sealing EPDM Fig. 7.8 Type: flow fitting Material: 1.4404, sealing EPDM 7.1.1 Transmitter with integrated transducer (compact version) The transverse hole in the measuring calotte must always be aligned parallel to the axis of the pipe, i.e.
7.1.2 Transmitter with separate transducer (wall-mounted version) The transmitter is suitable for wall-mounting. It is attached by 2 screws. The transverse hole in the measuring calotte must always be aligned parallel to the axis of the pipe, i.e. in the direction of flow of the medium. Deviations can have a distorting effect on the measured values (eddy formation).
2) Lines twisted to one another (black/red) 7.2.1 Auxiliary power connection According to EN 60204, Section 8, the LMIT 08 has a protective conductor connection in addition to the supply voltage connections X1/1,2. With respect to the wall-mounted transmitter, the potential equalization connection on the attachment flange can also be used to connect the protective conductor.
Fig. 7.14 Pos. Description PROFIBUS connection The PROFIBUS connection for the LMIT 08 is branched by a T-piece on the PROFIBUS line. The max. cable length of the individual line is 1 m. The shielding of the bus-connection wire inside the LMIT 08 has to be isolated.
Connection Fig. 8.1 Pos. Description LC display Contrast adjustment Programme storage Übergabestecker für CIP- bzw. PROFIBUS-Modul Operating buttons Pos. Symbol Function NEXT Selection of control levels Increase the displayed value Selection of a sub-menu Decrease the displayed value Selection of a sub-menu Acknowledgement of an input SEL/ Store the displayed value...
8.2.3 Display of uncompensated conductivity Fig. 8.4 The uncompensated conductivity, i.e. the conductivity actually measured can be displayed by pressing the SEL key. Also the initial current value for conductivity changes in relation to the actually measured NOTE conductivity value. 8.2.4 TCα...
8.4.1 Displays at the configuration level Fig. 8.9 The menu points are selected by the arrow keys and activated by the SEL key. 8.4.2 Interrogate Configuration 8.4.2.1 Measuring range interrogation Fig. 8.10 8.4.2.2 Current output assignment for conductivity resp. concentration Fig.
8.4.2.6 Unit number query Fig. 8.15 8.4.3 Change configuration 8.4.3.1 Code number protection Fig. 8.16 Access to the equipment configuration is authorized if the correct code number is entered. In order to disable the code number protection, it is also necessary to input the correct code number.
8.4.3.2 Change measuring range Fig. 8.18 If the "change measuring range" function is called up, the menu shown is displayed. Depending on whether a conductivity or product measuring range was set previously, this will be displayed on the second line. Use the arrow keys to choose between concentration and conductivity measurement.
8.4.3.5 Change assignment of the conductivity current output signal (zero point suppression and measuring range spread) LMIT 08 transmitter was designed in such a way that in the standard setting the starting and final value of the measuring range is assigned to the current output signal 0(4)... 20 mA.
- zero point: 0...50 °C - SPAN: 50...150 °C 8.4.3.7 Change device number Fig. 8.28 To every LMIT 08 you can assign an individual device number. Setting limits: 1...999 or --- if no device number should be assigned. 417101109_LMIT08.doc - 70 -...
8.4.4 Print out configuration values The RS232 printer interface is only suitable for printing out Fig. 8.29 parameterization data but not for on-line listing of conductivity and temperature values! In the event of a fault, check printer connection and printer status (online). Interface parameters: Baud rate 9600...
The assembly factor is a measurement unit that shows the influence the installation situation exerts on the constant of the measuring cell. The assembly factor enables the LMIT 08 calibration to be adapted to various installation situations. The table below shows the necessary assembly factors for tank installation and...
If a fault occurs: Fig. 8.39 8.5.5 Calibrate measuring cell with reference resistance LMIT 08 conductivity / temperature transmitter is calibrated exworks together with the measuring cell supplied. If recalibration is necessary following a replacement of the transducer, proceed as follows: •...
Page 75
Calibration of standard instruments: • Put the loop resistance with the indicated resistance value through the measuring hole. • Press SEL key to start Fig. 8.41 The calibration process is shown by a rotor in the display. After finish of a calibration for 1 second you get the message "calibration ok", and then the unit switches on to the next calibration value.
8.6.2.2 Concentration measuring range Fig. 8.46 Set PROFIBUS Abb. 8.47 For further information please see additional operating manual incl. disc with device data NOTE master file and software tools for Siemens PLC. 417101109_LMIT08.doc - 77 - Rev. 3-01.10...
The LMIT 08 can be tested electrically using a reference resistance (loop resistance) that is passed through the transverse hole of the measuring calotte. The loop resistance which is chosen according to the measuring range simulates a conductive liquid.
Page 79
Trouble Shooting FAULT SYMPTOM CAUSE/MALFUNCTIOM RECTIFICATION Check supply voltage or check Display does not light up, no supply "No display" whether miniature or fuse faulty if nec. voltage fuse is working replace fuse Air bubbles in installation site of Check measuring cell installation site measuring cell Insufficient pipe for air bubbles or inadequate pipe "Display unstable"...
CIP board 289152 PCB for PROFIBUS-DPV1 289156 PCB for PROFIBUS-PA 289158 Window printed LMIT 08 38910101 Standard measuring cells can not be combined with Pharma instruments and Pharma ATTENTION measuring cells do not match with standard instruments. 417101109_LMIT08.doc - 80 -...
Clamping ring set, 1.4404 415501231 Flow housing with flange connection, 1.4404 Housing with 2 connection systems (one side for connecting the LMIT 08 transducer, other side with detachable blank flange) DN 40 da 41 x 1,5 (AD x S) 415501261...
NaOH = 0 ... 5 % by weight / HNO = 0 ... 5 % by weight / H = 0 ... 5 % by weight Ecolab cleaning and disinfectant 0 … 5 % by weight product measuring ranges: Measuring range selection: via keyboard in unit (for "CIP-"version also via 24 V control signals) (for PROFIBUS...
Power supply Supply voltage: 24 V AC oder DC Tolerance: +/- 15 % Power consumption: approx. 6 VA Fuse protection: 5 x 20 mm miniature fuse 400 mA, slow-acting Electrical connection: plug-in screw terminals Ambient conditions Permitted ambient temperature: 0 … 50°C Influence of ambient temperature: <...
14.2 Diagrams 14.2.1 Specific conductivity Spec. conductivity æ of different electrolytic solutions according to concentration at 20 °C 1000 NaOH NaOH NaCl NaCl 0.6 0.8 1 Concentration in [weight %] 14.2.2 Temperature coefficients Spec. conductivity of different electrolytic solutions according to concentration at 20 °C. NaCl NaCl NaOH...
Relationship between temperature coefficients ar reference temperature 0 °C and 25 °C relative to 20 °C. TK- α (0 TK- α (0 TK- α (25 TK- α (25 Temperature coefficients α (20 °C) in [% / K] 14.2.3 Permitted operating pressure of the measuring cell Operating temperature in [°C] NOTE Only short-term operation is permitted in the critical range around 140 °C.
14.3 Connection diagram CIP 14.3.1 Example for signal input with relay 14.3.2 Example for signal with opto-coupler 417101109_LMIT08.doc - 87 - Rev. 3-01.10...
14.3.5 Connections in interconnections To avoid cross currents between conductivity and temperature output programmable NOTE controls, recording instruments and the like, of which the negative inputs are inter- connected, must be inserted at the final position in the current loop. Under no circumstances must 4...20 mA current interface cards be used, which deliver ATTENTION a voltage to supply two-wire transmitters.
14.5 Operating diagram display: see 8.4.2 change: see 8.4.3 print: see 8.4.4 default : see 8.4.5 back to display meas armature factor: see 8.5.2 TKα automatically: see 8.5.3 Temp. sensor: see 8.5.4 with resistor : see 8.5.5 back to display meas adjust adress: see range 1-4 see 8.6.2...
Instructions de sécurité Le transmetteur de conductivité et de température LMIT 08 peut uniquement être alimenté avec un courant continu ou alternatif de 24 V. Avant d’effectuer travaux de réparation, l’appareil doit être débranché et sans tension. Cet appareil de mesure a été construit et contrôler EN 61010-1, Partie 1, en ayant quitté...
Fig. 4.1 Le courant doit circuler de bas en haut. A ce propos, il convient de tourner le boîtier du LMIT 08 de façon que les connexions électriques montrent vers le bas (Fig. 4.1). Fig. 4.2 Dans le cas où...
Composants Version à clamp Fig. 5.1 Rep. Description Energie auxilaire 24 V AC/DC Sorties de courant 0(4)...20 mA cond. + temp. Transmetteur avec transducteur intégré (version compacte) pour installation dans conduites Fig. 5.2 Rep. Description Energie auxilaire 24 V AC/DC Sorties de courant 0(4)...20 mA cond.
Flanschversion Fig. 5.5 Rep. Description Energie auxilaire 24 V AC/DC Sorties de courant 0(4)...20 mA cond. + temp. Transmetteur avec transducteur intégré (version compacte) pour installation dans conduites Fig. 5.6 Rep. Description Energie auxilaire 24 V AC/DC Sorties de courant 0(4)...20 mA cond. + temp.
Description fonctionnelle Versions de l’appareil 6.1.1 Version BASE La version de BASE du transmetteur de conductivité et de température LMIT 08 contient toutes les fonctions nécessaires pour la surveillance et l’accroissement de milieux conducteurs. Versions de l’appareil : Appareil compact avec transducteur intégré pour installation dans conduites Appareil en version murale avec transducteur à...
6.1.2 Version CIP La version CIP du transmetteur de conductivité et de température LMIT 08 contient toutes les fonctions nécessaires pour la séparation des phases de milieux conducteurs. Versions de l’appareil : Appareil compact avec transducteur intégré pour installation dans conduites Appareil en version murale avec transducteur à...
6.1.4 Version Pharma La version « Pharma » du transmetteur de conductivité et de température LMIT 08 comprend toutes les fonctions du modèle de base et de la version CIP. Types d’appareil : Appareil compacte avec capteur de mesure intégré pour installation dans tuyauterie Appareil mural avec capteur de mesure séparé...
Fig. 7.7 Type: flasque à souder pour installation dans la paroi du réservoir Matériel: 1.4404, joint d’étanchéité EPDM Fig. 7.8 Type: ferrure de débit Matériel: 1.4404, joint d’étanchéité EPDM 7.1.1 Transmetteur avec transducteur intégré (version compacte) En principe, le perçage transversal de la calotte de mesure se fait parallèle-ment à l’axe du tube, c’est-à-dire en direction de courant du milieu.
7.1.2 Transmetteur avec trans-ducteur à part (version montage mural) Le transmetteur est approprié au montage mural. La fixation s’effectue avec deux vis. En principe, le perçage transversal de la calotte de mesure se fait parallèlement à l’axe du tube, c’est-à-dire en direction de courant du milieu. Déviations peuvent avoir pour effet l’altération des valeurs mésurées (formation de ATTENTION tourbillons).
Connexion énergie auxiliaire Selon EN 60204 section 8, outre les connexions pour la tension d’alimentation X1/1, 2, le LMIT 08 est pourvu d’une connexion pour le conducteur de protection. En cas d’utilisation d’un transmetteur en version montage mural, la connexion pour la liaison équipotentielle se trouvant au flasque de fixation, peut êgalement être utilisé...
Plan d’affectation des bornes, module PROFIBUS DPV1 et PA Fig. 7.14 Pos. Description Raccord PROFIBUS Le raccord PROFIBUS pour LMIT 08 est dérivé d’une pièce en T de la ligne PROFIBUS. Le câble de branche-ment a une longueur max. de 1 m.
Connexion Fig. 8.1 Pos. Description Afficheur-LC Réglage du contraste Mémoire des programmes Fiche de transfert pour module CIP ou PROFIBUS Touches de commande: Pos. Symbol Operation NEXT Sélection des niveaux de commande Augmenter la valeur affichée Sélection d’un submenu Diminuer la valeur affichée Sélection d’un submenu Confirmation de la saisie SEL/...
8.2.3 Affichage de la conductivité non équilibrée Fig. 8.4 En utilisant la touche SEL, la conductivité non équilibrée, c’est-à- dire la conductivité effectivement mesurée, peut être affichée. La valeur initiale du courant de la conductivité change aussi en fonction de la conductivité INDICATION effectivement mesurée.
8.4.1 Affichages dans le niveau de configuration Fig. 8.9 Les options sont sélectionnées à l’aide de touches flèches, et activées en utilisant la touche SEL. 8.4.2 Interroger configuration 8.4.2.1 Interrogation plage de mesure Fig. 8.10 8.4.2.2 Assignation des signaux de sortie de courant pour conductivité ou concentration Fig.
8.4.2.6 Interrogation numéro d’appareil Fig. 8.15 8.4.3 Modifier configuration 8.4.3.1 Protection numéro code L’accès à la configuration de l’appareil est donné, dès que Fig. 8.16 le numéro de code correct est introduit. Pour la désactivation de la protection du numéro de code il faut également entrer le numéro de code correct.
8.4.3.2 Modifier plage de mesure Fig. 8.18 Si l’application « Modifier plage de mesure » est appelée, le menu représenté est affiché. Les plages de mesure de conductivité ou du produit préalablement réglées sont affichées à la deuxième ligne. Les touches flèches servent à choisir entre mesurage de concentration ou de conductivité.
8.4.3.5 Modifier l’assignation du signal de sortie de courant de conductivité (suppression du zéro et écartement de la plage de mesure) Le transmetteur LMIT 08 a été construit de manière que, dans le réglage standard, la valeur initiale et finale est assignée au signal de sortie de courant 0(4)…20 mA.
8.4.3.5.2 En cas de mesurage de concentration Fig. 8.25 Limites de réglage : - Point neutre : 0…80 % de la plage de mesure - SPAN: 20...100 % de la plage de mesure 8.4.3.6 Modifier l’assignation du signal de sortie de courant de température (suppression du zéro et écartement de la plage de mesure) Fig.
8.4.4 Imprimer valeurs de configuration Tandis que l’interface d’imprimante RS232 permet l’impression Fig. 8.29 de données de configuration et de paramétrage, la constitution en ligne d’un listage imprimé de valeurs de conductivité et de température n’est pas possible. En cas de perturbations lors de l’impres-sion, il faut contrôler la connexion et l’état (en ligne) de l’imprimante.
8.4.5.1 Configuration usine • Plage de mesure de conductivité: 0..200 mS/cm • Température de référence: 20 °C • Tkα: 2,0 %K • Sortie de courant conductivité: 4...20 mA, correspond à 0...200 mS/cm • Sortie de courant température: 4...20 mA, correspond à 0...100 °C •...
Le facteur d’installation est une grandeur pour l’influence de l’endroit de montage sur la constante de cellule du capteur. Le facteur d’installation permet une adaptation du calibrage du LMIT 08 aux différentes situations de montage. Le tableau ci-dessous renvoie aux facteurs d’installation nécessaires pour l’installation dans un réservoir et pour les armatures d’écoulement qui font...
Calibrer le capteur avec résistance de référence Au moment de la fabrication à l’usine, le transmetteur de conductivité/température LMIT 08 a êté calibré avec le capteur faisant partie de la livraison. Si un nouveau calibrage est nécessaire après le remplacement du transducteur, il faut procéder comme suit :...
Page 117
Calibrage des appareils standard : • Conduire résistance de boucle avec la valeur de résistance indiquée à travers le trou de mesurage. • Appuyer sur la touche SEL pour démarrage Fig. 8.41 Un rotor montre le processus de calibrage sur l’afficheur. A la fin d’un processus de calibrage le signal «...
8.6.2.2 Plage de mesure concentration Fig. 8.46 Régler PROFIBUS Fig. 8.47 Pour informations complémentaires voir les instructions de service additionnelles avec INDICATION disquette du fichier maître des données de l’appareil et outils logiciel pour Siemens PLC . 417101109_LMIT08.doc - 119 - Rev.
NAMUR-NE21. Travaux d’entretien Les travaux d’entretien sur le LMIT 08 consistent essentiellement à contrôler à certaines intervalles, s’il existent des dépôts dans le trou de perçage transversal de la calotte de mesure, et à nettoyer le trou en cas de besoin.
Détection et élimination de défauts DEFAUT-SYMPTÔME CAUSE/MALFONCTIONNEMENT REMEDE Contrôler si tension d’alimentation Afficheur pas allumé, pas de tension est présente, ou vérifier si fusible « Pas d’affichage » d’alimentation ou consolidation pour courant faible est intact, défectueuse remplacer consolidation en cas de besoin Contrôler s’il existent des bulles Bulles d’air dans le lieu d’installation du...
289152 Platine PROFIBUS-DPV1 289156 Platine PROFIBUS-PA 289158 Fenêtre imprimée LMIT 08 38910101 Capteurs de mesure standard ne peuvent pas être combinés avec des appareils de ATTENTION version Pharma et les capteurs des appareils Pharma ne peuvent pas être utilisés dans des appareils standard.
Jeu anneau tendeur, 1.4404 415501231 Boîtier de débit avec raccord à flasque, 1.4404 Boîtier avec 2 systèmes de connexion (un côté pour raccorder le transducteur LMIT 08, l'autre avec collet blanc détachable) DN 40 da 41 x 1,5 (AD x S) 415501261...
= 0 ... 5 % par poids Plages de mesure des produits de 0 ... 5 % par poids netto-yage/désinfection de Ecolab: Sélection plage de mesure: à travers clavier dans l'appareil (avec version "CIP" même via signaux de commande au-dessus de 24 V) (avec version "PROFIBUS"...
Alimentation électrique Tension d'alimentation: 24 V AC our DC Tolérance: +/- 15 % Puissance absorbée: env. 6 VA Protection par fusible: 5 x 20 mm fusible pour courant faible 400 mA, lent Connexion électrique: bornes à vis enfichables Conditions ambiantes Température ambiante autorisée: 0 ...
14.2 Diagramme 14.2.1 Conductivité spécifique Conductivité spécifique æ de différentes solutions d'électrolytes en fonc-tion d'une concentration à 20 °C 1000 NaOH NaOH NaCl NaCl 0.6 0.8 1 Concentration en (poids %) 14.2.2 Coefficients de température Dépendance de concentration du coefficient de température avec de différentes solutions d'électrolytes (temp.
Dépendance de concentration du coefficient de température en cas des températures de référence 0 °C et 25 °C par rapport à 20 °C. TK- α (0 TK- α (0 TK- α (25 TK- α (25 Coefficient de température α en (20 °C) in [% / K] 14.2.3 Pression de service du capteur autorisée Température de service en (°C)
14.3 Schéma de raccordement CIP 14.3.1 Example pour entrée du signal avec relais 14.3.2 Exemple pour entrée du signal avec coupleur optoélectronique 417101109_LMIT08.doc - 129 - Rev. 3-01.10...
14.3.5 Connexions en interconnexions Afin d'éviter courants transversaux entre les sorties de courant de conductivité et de température, les commandes et enregistreurs programmables par mémorisation ou INDICATION choses semblables, dont les entrées négatives sont inter-connectées, doivent être insérés qù'à la fin du boucle de courant. Il ne faut absolument pas utiliser cartes d'interface de courant 4 ...