First set the measuring range upper and lower limit values via PCA or hCA (software or hardware cal-
ibration). See chap. 5.2 page 16.
Then select linearization programming:
cod4 = x6x (set the second digit of the parameter cod4 to 6). Attention: The remaining cod4 digits
must have the same values as before.
Enter the ten node values P10, P20, ... P100; store each with the P key. After entry of P100 the
measuring instrument again displays parameter cod4.
Now activate linearization: cod4 = x5x (set the second digit of cod4 to 5; exit with the P key. The
measuring instrument now displays the measurement value which corresponds to the selected linear-
ization.
Example: Programming the non-linear characteristic curve for thermocouple type S
(Pt10 %Rh/Pt)
Select the following values for offset and scaling factor:
OFSt = 0, SCAL = 1.0000 (see PCA, chap. 5.2 page 16).
Then proceed with linearization programming as described above and enter the following 10 node
values:
P10 = 000
P40 = 828
P20 = 255
P50 = 998
P30 = 649
P60 = 1161
Next, activate linearization: cod4 = x5x (set the second digit of cod4 to 5. Attention: The remaining
cod4 digits must have the same values as before. Exit with the P key. The measuring instrument now
displays the measurement value which corresponds to the type S thermocouple characteristic curve.
5.4
Activation of Automatic Tare
Set the first digit of cod3 to 3 (cod3 = 3xx).
Attention: The remaining cod3 digits must have the same values as before. Press the P key repeat-
edly, until the instrument returns to the normal operating mode.
5.5
Programming and Activation of a Display per cosϕ
First deactivate display of cosϕ: cod3 = 0xx (set the first digit of parameter cod3 to 0 = dark). Atten-
tion: The remaining cod3 digits must have the same values as before. Press the P key repeatedly,
until the instrument returns to the normal operating mode. Then select either hCA (hardware calibra-
tion) or PCA (software calibration).
Set the display range according to the angular degree of cosϕ with a resolution of 0.01 degrees.
Example: Range cosϕ = – 0.5 ... 1 ... 0.5
Set display range = – 60.00 ... 00.00 ... 60.00
Now activate the display for cosϕ: cod3 = 1xx (set the first digit of cod3 to 1). Attention: The
remaining cod3 digits must have the same values as before.
Press the P key repeatedly, until the instrument returns to the normal operating mode. The measur-
ing instrument now displays the measurement value per cosϕ.
18
P70 = 1319
P80 = 1477
P90 = 1638
P100 =1807