Shockwave Medical 825D Manuel D'utilisation page 27

Lithoplasty générateur
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 74
Observation
Unit will not turn on
Battery will not charge
(BATTERY CHARGING indicator is off)
Low BATTERY CAPACITY indicated
when the battery symbol is Empty
System Error displayed and red light
around POWER ON/OFF button.
Catheter Error Displayed
THERAPY STATUS indicator on
Generator front panel or on the
Connector Cable does not turn on
Generator will not activate (THERAPY
STATUS is off)
Generator will not activate (THERAPY
STATUS is yellow)
THERAPY STATUS changed from green
to yellow
THERAPY ON/OFF button on the
Connector Cable is pressed, but
Generator does not deliver pulses
LBL 60780 rev. B
Possible Cause
Battery requires charge
Disconnected cable
Battery requires charge
Built in tests have detected a malfunction
within the Generator
Catheter connection has not been made
properly or has become disconnected.
Lithoplasty Catheter has reached its
useful life or is defective
Connector Cable has reached its useful
life
Valid Lithoplasty Catheter is not
connected
Valid Lithoplasty Catheter is not
connected
THERAPY ON/OFF button is pressed
System has automatically paused therapy
(see Section 4.0 Step 9)
Generator cannot be activated
(THERAPY STATUS indicator is off)
Generator has not yet been activated
(THERAPY STATUS indicator is yellow)
System has automatically paused therapy
(THERAPY STATUS indicator is yellow, see
Section 4.0 Step 9)
Lithoplasty Catheter or Connector Cable
is faulty
(THERAPY STATUS indicator is green)
Corrective Action
Connect Charger Module to the
Generator and to AC Mains.
Allow the Generator to charge at least
twelve (12) hours prior to use
Connect Charger Module to the
Generator and also to AC Mains.
NOTE: two cable connections are
required.
Connect Charger Module to the
Generator and to AC Mains.
Allow the Generator to charge at least
twelve hours prior to use
Turn Generator off, wait one second and
turn the Generator on again.
NOTE: If SYS fault cannot be resolved,
remove Generator from service and
contact your Shockwave Medical
representative.
Make sure Connector Cable is connected
to Generator
Make sure valid Lithoplasty Catheter is
connected to Connector Cable
Replace Lithoplasty Catheter
Replace Connector Cable
Make sure Connector Cable is connected
to Generator
Make sure Lithoplasty Catheter is
connected to Connector Cable
Replace Lithoplasty Catheter
See THERAPY STATUS indicator
troubleshooting step above
Release THERAPY ON/OFF button, try
again
THERAPY STATUS should indicate green
again, within the pause period specified
in the Lithoplasty Catheter IFU
If error message(s) appear, see
troubleshooting guide above
Press THERAPY ON/OFF button once
(THERAPY STATUS should turn green)
THERAPY STATUS should indicate green
again, within the pause period specified
in the Lithoplasty Catheter IFU
Replace Lithoplasty Catheter
Replace Connector Cable
27 |
P a g e
EN