Instructions specific to hazardous area installations
(Reference European ATEX Directive 94/9/EC,
The following instructions apply to equipment covered by certificate number SIRA
13ATEX5017X.
1.
The equipment may be used with flammable gases and vapors with the Apparatus Group
II and with Temperature class T5.
2.
The equipment is certified for use in an ambient temperature range of -20 °C to +60 °C
3.
The equipment has not been assessed as a safety related device as referred to by EC
Directive 94/9/EC Annex II, clause 1.5.
4.
The certificate has a 'X' suffix, which indicates that special conditions of installation and
use apply. Those installing or inspecting this equipment must have access to the contents
of the certificate.
5.
The apparatus shall only be supplied from a circuit containing a suitably rated fuse having
a breaking capacity of at least 1500A.
6.
Installation and inspection of this equipment shall be carried out by suitably trained
personnel in accordance to the applicable code of practice (EN 60079-14 and EN 60079-17
within Europe).
7.
Repair of this equipment shall be carried out by suitably trained personnel in accordance
with the applicable code of practice (e.g. EN60079-19 within Europe).
8.
Components to be incorporated into or used as replacements in the equipment shall be
fitted by suitably trained personnel in accordance with the manufacturer's
documentation.
9.
The certification of this equipment relies upon the following materials used in its
construction:
Norton Performance Plastics Chemgrip bonding compound
ETFE housing material
PVDF housing material
Dow-Corning 3-4207 encapsulant
Royal EPOCAP 42174/52174
Cotronics Durapot 864 encapsulant
If the equipment is likely to come into contact with aggressive substances, such as solvents
that may affect polymeric materials, it is the responsibility of the user to take suitable
precautions to prevent the equipment from being adversely affected. It is important to perform
routine inspections and establish from the materials data sheet that the transducer is resistant
to chemicals used in the application.