Notizen / Notes / Note
Required software
The EBI16 is supported:
- Winlog.med
- Winlog.validation
If possible, use the current version.
Logger and interface
- To program or to start recording the
EBI16 must be placed on the
interface.
- To read out the data of the EBI16 it
must be placed on the interface.
- Make sure that the case of the EBI16
is not hotter than +70°C. It could
damage the housing of the interface
otherwise
Cool down times
- To prevent damage to the housing of
the interface, the EBI16 must cool
down for approx. 7 minutes before it
can be placed on the interface.
- To reach the start condition of
+15°C ... +35°C -for programming-
after a cycle with 134°C the EBI16
must cool down at room temperature
for approx. 45 to 60 minutes.
03/2021 BA80000def05_EBI 16
Français
Logiciel requis
L'EBI 16 peut être utilisé avec :
V 3.71
- Winlog.med
V 3.71
- Winlog.validation
Si
actuelle.
Enregistreur et interface
- Pour
l'enregistrement, l'EBI 16 doit être
placé sur l'interface
- Pour lire les données de l'EBI 16, il
faut également le placer sur l'interface
- Attention à ce que la température du
boîtier de l'EBI 16 n'excède pas
+70°C. Cela pourrait endommager le
boîtier le l'interface
Durée de refroidissement
Afin d'éviter tout dommage au niveau
-
du boîtier de l'interface, l'EBI 16 doit
refroidir environ 7 minutes avant d'être
positionnés sur l'interface.
- Lors d'un cycle à +134°C, pour
atteindre la condition de départ entre
15°C et 35°C, il faut laisser reposer
l'EBI 16 environ 45mn à 60mn à
température ambiante.
possible,
utilisez
programmer
V3.71
V3.71
la
version
ou
démarrer
21