Guidance and manufacturer's declaration – electromagnetic immunity
The CapsoCam Plus is intended for use in the electromagnetic environment specified below.
The customer or the user of the CapsoCam Plus should assure that it is used in such an environment.
Immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
Document Number: IFU-2320 Rev: H Revision Date: 8/28/2019
(Table 3)
IEC 60601
Compliance level
Test Level
Not Applicable
20 V/m
Electromagnetic environment –
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the EQUIPMENT, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 0.18√P 80 MHz to 800 MHz
d = 0.35√P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey
, should be less than the
a
compliance level in each frequency
range.
b
11