3. Troubleshooting
Problem
Sticking AgCl precipitates.
Insufficient cleaning.
Rinsing solution is dripping.
6.5613.000 Titration Equipment
Corrective action
• Add anti-flocculating agent (e.g.
Polyviol®).
• Feed the 703 Ti Stand via separate
power adapter (see page 26).
• Place the inserts (electrode, buret
tip) directly before the rinsing noz-
zles.
• Work with only 1 rinsing nozzle.
Place the inserts (electrode, buret
tip) as closely as possible to this
rinsing nozzle.
3. Troubleshooting
24