Measurement Data - Citizen CH-617 Mode D'emploi

Masquer les pouces Voir aussi pour CH-617:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

c. Press"
seconds. Afterward, displays the updated time. The
measuring blood pressure symbol will flash on the display
and then the air pressure will automatically pump up to
195 mmHg. If blood pressure tends to be on the high
level, the air pressure will automatically increase to the
required level for measurement.
* Do not move or talk in the midst of taking blood pressure
measurement.
d. After the air pressure is increased, it will slowly decrease.
When the pulse rate is detected, HEART RATE symbol
will start flashing.
*Note!
• If the cuff does not stop inflating, remove the cuff at once.
• To interrupt the measurement, you may press the
The cuff will deflate immediately after a button is pressed.
e. After taking blood pressure measurement, the Systolic rate,
Diastolic rate and pulse rate will be on the display for 1 minute.
D. Storing, recalling and erasing measurement data:
a. Storing data:
After each blood pressure measurement, the Systolic rate,
diastolic rate, heart rate and the time & date of specific day
will be automatically stored. The memory holds the latest
99 measurement data. If more than 99 measurements, the
memory will automatically clear out the earliest data.
"button. All elements display '888' in 3
13
button.