Power-up options:
Press button while turning meter on from OFF position.
Lock button: display of the software version.
Store button: reset all stored data.
Blue button: disable auto power off.
Battery and Fuse Replacement
Refer to the following figure to replace fuse and batteries.
Testing the Fuse
COMP
TEST
100V
0
50V
Ω
LPF
V
OFF
Ω
>600V
COM
FUSED
61-797
TEST
RECALL
PI/DAR
STORE
LOCK
PASS
250V
500V
1000V
V
INSULATION
CAT. IV
600V
CAUTION:
• Use only a fuse with the
amperage, interrupt, voltage,
and speed rating specified.
• Fuse rating : Fast, 315mA,
1000V, Min Interrupt
Rating 10000A.
• Replace the batteries as
soon as the low battery
indicator "
to avoid false readings.
• When the battery is too low
for reliable operation, the meter
displays "bAtt." The meter will
not operate at all until the battery
is replaced.
• 1.5V x 4 alkaline batteries.
TEST
• Remove the test leads before
testing the fuse.
• If the display is reading FUSE,
the fuse is bad and should be
replaced.
14
" appears,