Electrical Specifications; Wiring Diagram; Agency Approvals; Leakage Current - Welch Allyn LS200 Manuel De L'utilisateur

Appareil d'éclairage articulé
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ELECTRICAL SPECIFICATIONS

Wiring Diagram

A
B
A
HOSPITAL GRADE POWER CORD
B
FUSES
C
TRANSFORMER - STEP DOWN,
ISOLATED
Input/Output
Model 44200: 120 Vac, 60 Hz 0.70 A
Model 44202: 230 Vac, 50 Hz 0.35 A (Europe, except
Models 44204, 44206: 240 Vac, 50 Hz 0.35 A (Australia, UK)

Agency Approvals

Model 44200, listed to: UL 544 — Class 1 Medical Equipment
Models 44202, 44204,
44206, listed to: IEC 601-1 — Class 1 Type B Medical

Leakage Current

C
Input:
(U.S.A./Canada/South America/Japan)
United Kingdom)
Output: 12.0 Vac @ 5.0 A (three 20-watt lamps)
CSA 22.2-125 Class 2G Medical Equipment
Equipment
The CE mark on this product indicates it has
been tested to and conforms with the provi-
sions noted within the 89/336/EEC Electro-
magnetic Compatibility Directive.
Less than 50 µA from any metal part
D
D
ON-OFF SWITCH
E
LAMPS
13
E

Publicité

Table des Matières
loading

Produits Connexes pour Welch Allyn LS200

Table des Matières