11. Replace bezel. Be certain to align the three smooth "windows"
in bezel with lamps and tabs on edge of bezel with grooves in
white housing (see J).
12. Replace handle and rotate clockwise until secure (see J).

FUSE REPLACEMENT

There are two replaceable fuses located in the fuse drawer of the
transformer housing. If either one of these blows, the unit will not light.
Replace only with Welch Allyn part number 236706-3907 for model
44200 or 236706-3114 for models 44202, 44204 and 44206.
1. Turn the power switch to the OFF position and unplug the power
cord from the electrical outlet and transformer.
2. Using a small (1/8") screwdriver, remove the fuse drawer (a)
from the transformer housing by releasing the spring loaded snap
connection (b) on both sides of the component.
3. Pull out fuse drawer.
4. Remove and replace the blown fuse(s). There is no required
orientation of the fuses.
5. Reinsert the drawer by pressing it into the connector until it
snaps into place.
I
(a)
FUSE
DRAWER
ALIGN SMOOTH
WINDOWS IN
BEZEL WITH
LAMPS
TURN
CLOCKWISE
J
(b)
SNAP
CONNECTION