7.1 SIDERAIL COVERS
WARNING: Only use M944-03 siderail covers for M9 Transfer Stretchers (Horizontal siderails).
•
Inner: Bi-elastic foam
•
Cover: Polyurethane coated stretch knit fabric
To fit:
1. Pull cover over the raised siderails. Ensure the long strap and clips are on the inside of the siderail.
2. Fasten the two buckle clips together from under the bottom siderail tube. Pull straps to tighten.
NOTE:
Only use siderail covers with the siderails in the raised position.
NOTE:
Please see Siderail Covers Instructions for use (M799-69) for further information.
7.2 MATTRESSES
The mattress best suited to the M9 Transfer stretcher is the PREMA Stretcher.
The recommended mattress size for the M9 Transfer stretcher is 2000mm long x 700mm wide x 100mm thick.
The maximum mattress thickness is 160mm.
NOTE:
The mattress cover can be removed and machine washed (95°C max). For cleaning and disinfecting
information please visit www.howardwrightcares.com
NOTE:
Please see PREMA Stretcher Mattress Instructions for use (M799-72) for further information.
7.3
ORTHOPAEDIC FRAME
WARNING: The safe working load of the frame is 100kg and 30kg (weights).
WARNING: It is recommended two people assemble the frame.
WARNING: Apply brakes and lower the stretcher to the minimum height before starting frame assembly.
WARNING: Ensure all clamps are tight before applying any load to the frame.
WARNING: Do not use a monitor tray with the orthopaedic frame.
WARNING: Use knots suitable for the application e.g. surgeon's loop.
The frame plugs into the
four corner
NOTE:
Please see Orthopaedic Frame Instructions for use (M799-67)
for assembly instructions and further information.
accessory sockets on the deck.
M9 Transfer Instructions For Use (M999-190, Version 1, 2022-04)
7. ACCESSORIES
Figure 26. Orthopaedic frame
21