PROBLEMS
Excessive noise
The unit turns on but it does not suck
The vacuum value cannot be adjusted
The protection fuse is activated any time
the device is turned on
The vacuum gauge does not work
Note: if you experience faults or malfunctioning problems different to those listed above, always and exclusively contact authorised
assistance centres.
Piston electric-compressor, with thermal protection.
Fuse
Supply voltage
Adjustable vacuum level
Air flow
Dimensions
Weight
Noise level
Operation cycle
Class of risk according to the 93/42/EEC directive
Operating Temperature/Humidity
Storage and Transport Temperature/Humidity
Operating-storage air pressure
IMPORTANT INFORMATION REGARDING ELECTRO MAGNETIC COMPATIBILITY (EMC)
Electronic devices (PC's, mobile phones, televisions, etc) that generate electromagnetic interference can affect the correct functioning of
medical devices. This situation can lead to incorrect functioning creating unsafe conditions.
In order to show compliance with the Directives on Electromagnetic Compatibility (EMC), all the tests introduced by the standard
EN60601-1-2:2015 have been implemented.
This standard defines the levels of immunity to electromagnetic interference as well as maximum levels of electromagnetic emissions
for medical devices: 3A HEALTH CARE products comply with the levels specified in the standard. However, when the device is in use,
it is best practise not to use mobile phones or other electronic devices which generate strong electromagnetic fields near the medical
device. This may create dangerous conditions that put the successful outcome of treatment at risk. On the basis of experience and tests
PROBLEMS, CAUSES AND SOLUTIONS
POSSIBLE CAUSES
Damaged pump or obstructions in the internal
suction duct
- Damaged pump
- Vacuum regulator fully open. Connection
tubes disconnected and/or badly connected,
broken connection tubes. Container not in a
vertical position, full, or defective overflow
valve. Possible blockage of the hydraulic cir-
cuit inside the unit.
Damage to the internal hydraulic circuit or
obstruction of the connection tubes to the
aspiration unit
Pump damaged or in short circuit
Liquid penetrating the pneumatic circuit.
TECHNICAL SPECIFICATIONS
Single pump version
Low flow, high vacuum device
230V ~ 50Hz 150VA
approx. 15 l/min.
2,7 kg approx.
approx. 63 dBA
10
Send to the customer care service
- Send to the customer care service.
- Check the position of the vacuum regula-
tor. Check the connections and the integri-
ty of the tubes. Position the container in a
vertical position, check the overflow valve
(blocked) and/or replace the silicon tubes.
Send to the customer care service
Send to the customer care service
Send to the customer care service
Double pump version
High flow, high vacuum device
T1, 6A - 250V
0 ÷ -0,85 bar (85 kPa)
196 x 357 x 185(H) mm
Continuous use
IIa
+10° C to +40° C/10% to 95% RH
–25° C to +70° C / 10% to 95% RH
min. 690 hPa - max 1060 hPa
SOLUTIONS
230V ~ 50/60Hz 100VA
approx. 22 l/min.
3,4 kg approx.
approx. 65 dBA