The ASPEED 2 aspirator is a professional device suitable for home and outpatient use, specific for secretion-aspiration. It is equipped
with Vacuum regulator 1, vacuum gauge 4, 1.000 ml vessels. with protection device, for the liquid intake on the suction pump, that
interrupts the suction flow. It is lubrication-free, handy, easy to use, reliable, sturdy and quiet.
The ASPEED 2 aspirator is supplied with the following 3A accessories: 1.000 ml vessel with protection device, power supply cable,
connection tube in sterilisable silicon Ø 6x12 short, connection tube in sterilisable silicon Ø 6x12 long, sterile and single use
cannula, sterile and single use manual regulator, single use bag, single use antibacterial filter.
Note: Use original 3A accessories ONLY.
This is a medical device and it must be used only upon medical prescription. It must be used as described in this
instruction manual. The user must read and understand in full the information regarding the use and maintenance
of the unit. For any question, do not hesitate to contact your preferred retailer.
MICROBIAL CONTAMINATION: due to pathologies with risk of microbial infection and contamination a personal use
of the accessories is recommended (ask your physician).
The manufacturer does anything possible to provide a product equipped with the highest quality and safety features, although, being it an electric
device, some fundamental safety directions must be observed:
• Kids and non self-sufficient people must always use the device under strict supervision of a competent adult in his full-capacity and who has
read and understood this manual. Ensure that children do not use the device unsupervised some parts are small enough to be swallowed.
• The device must always be used by specifically trained staff who have read this manual.
• This device must only be used for its intended purpose: aspirator for outpatient and domestic use; any other use is considered improper and
dangerous and the manufacturer cannot be held responsible for any consequence of misuse.
• Never use adapters for voltages different from the voltage indicated on the data label on the bottom of the device.
• Keep the cable away from hot surfaces.
• This device is not suitable for use in presence of anaesthetic blend inflammable with air, oxygen, or nitrous oxide.
Never handle the power supply cable plug with wet hands, and never use the device in the bath or in the shower.
• Never leave the device near water, never immerse it in any liquid, never wet it, if it gets wet immediately unplug it from the power supply socket
before touching it. Do not use the device if the plug or the power supply cable are worn or wet (immediately send it to your preferred retailer).
• The device is not watertight.
• Always unplug the power supply cable immediately after use.
• Maintenance and/or repair must be carried out only by authorised personnel. Non-authorised repairs void the warranty.
• Ensure that the connections and vessel closing are performed carefully to prevent suction losses.
• Do not turn over the vessel while it is connected to the device in function, since the liquid would come into contact with the hydro phobic
antibacterial filter, immediately blocking suction; in case this happens, empty the vessel and replace the antibacterial filter.
• In the event of aspiration without the container and/or antibacterial filter, or if it is suspected that solid and liquid substances have entered the
aspiration circuit, it is necessary to send the device to the customer care service.
• In view of their length, the power cord and connection hose could constitute a strangulation hazard.
• The device is equipped with a safety fuse, mounted inside the device, in case you wish to replace it, unplug the device before pro-
ceeding.
• The pipe and the manual control of the flow are sterile and single-use products: they must be replaced after each application. Check the
expiration date on the original packaging of the pipe and the suction flow manual control and verify the integrity of the sterile packaging.
• The antibacterial filter is a single-use device and it must be replaced after each application.
• The lifetime of the device itself is 5 years, and that of the container and silicone tubes 1 year or 30 sterilisation cycles. However, the lifetime
may vary depending on the operating environment.
• Before each use always perform the cleaning and disinfecting operations as mentioned in the paragraph "CLEANING AND DISINFECTING
OPERATIONS" of this instructions manual.
Before each use, verify that all the accessories are perfectly clean in accordance with the instructions of the section "CLEANING AND
DISINFECTING OPERATIONS".
1. Connect the device as shown in figure 1 (page 2).
2. With the vacuum regulator 6 you can set the depression value (bar). Rotate the knob towards «+» to increase the vacuum and rotate
it towards «-» to decrease it, the values can be read on the vacuum gauge 7.
3. Turn the device on by putting the switch on the «I» position (ON) 3.
4. After the application turn off the device and unplug the cable form the power supply socket and carry out the cleaning and disinfecting
EN

IMPORTANT WARNINGS

INSTRUCTION FOR USE
8