ZP 0
• Use the key
ESC to exit from the configuration and go back to the menu LF4)
LF 5
• Use the keys
key
APPROVAL CONFIRMITY
DISPLAY
nonE Version not approved
LF 6
DON'T MODIFY, IN NO WAY, THIS PARAMETER. IT MUST ALWAYS BE CONFIGURED AT
nonE.
!
The change of this parameter implies the blocking of some functionalities.
GRAVITY
LF 7
*Introduce the gravity of your zone before doing the first calibration.
*Introduce the gravity of destination after doing the calibration.
*The value of the gravity will be denied if it is bigger than 9.83217 ( gravity of the pole) or inferior to 9.78031 ( gravity of the
equator).
Value of factory: 9.8035
DISPLAY
• Press the key
LF 7
again automatically.
-00-
• The screen is going to show the number of pre-calibration during one second.
9.8035
• Press the key
9.8035
• Use the keys
gravity ( press ESC to exit from the configuration and go back to the menu LF7)
INITIAL ZERO
LF 8
SEtZ Y resets of the point of zero every time that the scale is begun again.
SEtZ n resets of the point zero OFF
DISPLAY
• Press the key
LF 8
automatically.
and the key
,
and then the key
to exit from the menu and the scale is going to start automatically.
LF 6
DESCRIPTION AND SEQUENCE OF USE
DESCRIPTION AND SEQUENCE OF USE
to continue or the key
to continue.
,
and
DESCRIPTION AND SEQUENCE OF USE
to start or the key
to select how many divisions will show when in zero ( press
to continue with other adjustment or press the
to exit from the menu and the scale will start
0 ~ 9
,
and then the key
to exit from the menu and the scale will start again
USER'S MANUAL BR95
to introduce the value of
EN