Unit Range/ Multi Range/ Multi Interval; Parameters; Parameters Configuration; Internal Counting (A/D) - Giropes Baxtran BR95 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

The user can configure the superior and inferior limits of the sample placed on the platform.
The display will show if the sample is lower to the inferior limit Lo, above the superior limit Hi or in the zone between the two limits OK
The user can configure when he wants to make the alarm ring and the stability needed to make it happen.
All the procedure is described in the section UF-2.
If the user wants to define the limits in the normal weighing mode and wants to use the limits in piece counting mode, he must define the new
limits for this mode, when it changes to weight mode again, the user will recover the limits he already has. The same happens otherwise.
6.4 UNIT RANGE / MULTI-RANGE / MULTI-INTERVAL
The indicator can be configured with an only range, a maximum weight and a value of step. It can also be configured as multi range or multi interval,
in such cases there is a maximum weight and two steps.
From 0 to the medium weight of these maximum weight it is used the value of the chosen step ( step 1) and from the half to the maximum weight
it is used the next value in the step ( step 2).
The screen indicators R1 and R2 point out the range which the user is using at every moment.
In the multi interval mode the weight increases, in the range use it used step 1, when the user goes to range 2 the step 2 is used.
When the weight decreases and the user go back to range 1, the step 1 is used again. On the contrary, in mode multi range, when the weight
decreases and the device go back to range 1, the device continues using step 2 until it reaches 0.
In the section LF2 of the technical parameters the user can choose the range mode.

7. PARAMETERS

8. PARAMETERS CONFIGURATION

8.1

INTERNAL COUNTING (A/D)

8.2

CONFIGURATION OF THE WEIGHT LIMITS (SUPERIOR AND INFERIOR)

8.3

AUTO SWITCH OFF

8.4

CONFIGURATION OF THE ILLUMINATION OF THE DISPLAY

8.5

HOLD FUNCTION

8.6
RS-232 DATA EXIT
8.7

CONFIGURATION OF THE SPEED OF THE AD CONVERTER

8.8

BLIND

8.9

CONFIGURATION OF THE GRAVITY

8.10
CONFIGURATION OF THE WEIGHING FILTER
8.11
CONFIGURATION OF THE STABILITY FILTER
8.12
CONFIGURATION OF THE INITIAL ZERO
To access the configuration of parameters when the screen is in zero, the user must press at the same time the keys
Press the key
or
To go back to the previous mode press the key
PARAMETER
DESCRIPTION
Internal calculation (A/D)
UF-1
Limit Configuration of weight ( superior and inferior)
UF-2
auto auto off
UF-3
Back illumination of display
UF-4
Four modes of hold
UF-5
RS-232 Output ( PC/PRINT)
UF-6
Configuration of the speed of the converter (A/D)
UF-7
blind
UF-8
Configuration of gravity
UF-9
o choose the desired character (UF-1 ~ UF-11), for confirm the changes done, press
.
USER'S MANUAL BR95
and
.
EN
35

Publicité

Table des Matières
loading

Table des Matières