33
peratIng the unIt
Set the parameter
TENS mode, in user programs (P1, P2), the user can set the pulse
DE
rate, pulse width and treatment time by press ["SET" button (F). The
user LcD will be display the current parameter which you want to
set like following:
EN
FR
IT
EMS mode, in user programs (S, A), the user can set the pulse rate,
ES
pulse width, treatment time, ramp time, working time and rest time
by press "SET" button (F). The LcD will be display the current para-
meter which you want to set like following:
NL
RU
PL
SE
FI
And then press "
" (G) or "
After the parameter is set, press the "SET" button (F) again to set
next one.
Remark: The set range of pulse rate, pulse width, treatment time,
ramp time, working time and rest time in user program, please refer
to section programs (page 13-14). IF you want to restore the user
programs to factory reset, you can achieve by press the "
button and turn the device on at the off state mode.
Adjust the intensity and Start to treatment:
There are two intensity adjustment knobs (C) for each channel on the
side of the unit. By turning the controls up, the output intensity will
be increased, when the output intensity above 1mA, the device will
be started to work. The maximum output intensity is 100mA.
Remark:
The Promed EMT-6 has special backlight function:
a) In the normal state the backlight is green;
b) If the output intensity is over 65mA, the backlight will be changed
blue and the user should be taken care.
c) If the electrode pad connection in wrong way or not connect the
electrodes, when the output intensity over 10mA, the backlight
will change to red and flash, and the intensity will be reseted
automatically.
" (H) button to set the parameter.
" (G)