EN OPERATION
5. Use the access key to remove the front
access panel (machines serial number
ORB3001577 and below)
6. Insert blank USB drive into the USB port.
(ORB3000000 − ORB3001577)
NOTE: Do not use the BLUE Orbio USB flash
drive on machines serial number ORB3001578
and above or machines updated to the display
with front USB ports.
(ORB3001578−
7. Press pH icon on touchscreen.
16
(ORB3001578−
)
)
8. Confirm new pH probe installation.
9. Verify correct installation on the Wear
Component Status Indicator Panel. All
segments on the progress bar will display as
green indicating a new pH probe.
GREEN
10.Remove USB drive from the USB port.
11. Reinstall the front access cover (if needed).
12.Reinstall the top cover.
13.Update product label.
Place the MultiMicro Directions for Use label
included in the pH probe service kit over the
existing label if the revision level number on the
replacement label is greater. The label is located
on the lower right hand side of the os3 generator.
Note: The pH probe service kit may contain more
than one MultiMicro Directions for Use label.
Select the appropriate label by country.
pH
Revision #
OS3 DM20036 (1−2018)