The heater in the vacuum forming machine may absorb humidity or moisture when stored for a
period of time. This may penetrate to the internal connectors and cause the unit to short circuit
your GFCI outlet. If this occurs, we recommend using the machine on a non-GFCI outlet for
approximately twelve cycles. This will enable the moisture to dissipate.
The situation is rare and is not cause for an alarm. Should you wish to discuss it further, please
contact the Customer Service Center at: (800) 883-8733.
The Essix® Machine is intended for indoor use in a standard dental laboratory environment.
Ambient operating temperature:
Relative Humidity:
Altitude:
Installation Category:
Pollution Degree:
Main supply fluctuations not to exceed ± 10% of the nominal voltage
THE ESSIX® MACHINE SYMBOL KEY
Fuse
Ground
THE ESSIX® MACHINE TECHNICAL SPECIFICATIONS
Model #85000
Number 101
115 V
50/60 Hz
1270 W
115 VOLT
To Attached Cord
GREEN
Fuse-
15amp
TO FRAME
125volt
Black
WHITE
Model #85220
Number 202
230 V
50/60 Hz
1475 W
110 Volt
White
HEATER-110v
BLACK
BLACK
LIGHT-110v
BLACK
BLACK
M
MOTOR-110v
GREEN
BLUE
TO FRAME
5-40C.
Max 80%
Maximum altitude 2,000 m
nominally 11
nominally 2
On
Off
TO FRAME
To Attached Cord
220Volt
Fuse-
6.3amp
250volt
Fast Blo
WHITE
BLACK
FILTER
BROWN
WHITE
HEATER-220v
BLACK
BLACK
LIGHT-220v
BLACK
M
MOTOR-220v
Attention
Vacuum
Model # 85220-CE
Number 202CE
230 V
50/60 Hz
1475 W
230 VOLT
To Attached Cord
GREEN
GREEN
220Volt
Fuse-
6.3amp
250volt
WHITE
BLACK
BLACK
230 VOLT
CE MARKED
UNIT WITH
FILTER
Heater
HEATER-220v
BLACK
LIGHT-220v
M
MOTOR-220v
5