2.3.3 Storing results
Store
Code-No.:
_ _ _ _ _ _
0
1
0
0 0
Stored!
Storage: 900
free records left
Storage:
only 29 free records left
2.3.4 Printing results
If a printer is installed and switched on, it is possible to print out the test results (without
saving before).
F3
2.3.5 User calibration
2.3.5.1 When to calibrate
The turbidimeter was calibrated at the factory with Formazin Primary Standard and does not
require user calibration before use.
As the electronic and optical design of this instrument provides long-term stability the need
42
Press [STORE] during the test result is displayed.
The display shows:
•
We advise you to enter a numeric code (up to 6 digits
6
long). (A code-No. can contain references to the ope-
rator or the sampling location.)
After entering confirm with [ ] key.
•
If a code number is not necessary confirm by pressing
[ ] directly. (The assignment for the Code-No. is then
0 automatically.)
The data set is stored with date, time, code-No. and test
result.
The display shows:
The test result is then shown again.
Note:
The display shows the number of free data sets.
If there are less than 30 data sets free the display shows:
Clear the memory as soon as possible (see chapter 2.4.4
Delete stored results). If memory capacity is used up it
would be impo ssible to save additional test results.
Press [F3] key.
The entire data set is printed with date, time, code-No.
and test result.
Printing example:
2014-04-01 14:53:09
Test No.:1
Code-No.:1
2,13 NTU *
The Test No. is an internal number that is set automatically
if a test result is stored. It appears only at the print out.
TB300 IR 6 03/2022