Voltage dips
Voltage
interruptions
IEC 61000-4-11
S5 Apex Locator is intended for use in the electromagnetic environment specified below;
The customer or the user of S5 Apex Locator should assure that it is used in such an environment.
Immunity test
Rated power
frequency
magnetic fields
IEC 61000-4-8
Note:
• UT is the a.c. mains voltage prior to application of the test level.
Conducted
3 Vrms
disturbances
inducted by RF
150 kHz to 80
fields
MHz
6 Vrms in ISM
bands 150 kHz to
80 MHz
IEC 61000-4-6
80% AM at 1 kHz
Radiated RF
10 V/m
0% UT; 0,5 cycle
0% UT; 1 cycle
and
70% UT;
25/30 cycles
0% UT;
250/300 cycles
Guidance and manufacturer's declaration:
electromagnetic immunity - S5 Apex Locator
Immunity test
30 A/m
50 or 60 Hz
Not Applicable
Immunity test
30 A/m
50 or 60 Hz
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM bands
150 kHz to 80 MHz
80% AM at 1 kHz
10 V/m
23
Not Applicable
Immunity test
Power frequency magnetic
fields should be at levels
characteristic of a typical public
low-voltage power supply
network that supplies buildings
used for domestic purposes,
commercial or hospital, clinic
environment.
Portable and mobile RF
communications equipment
should be used no closer to any
part of S5 Apex Locator ,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1.17√P
d = 1.17√P
80 MHz t o 800
MHz