8. Demo mode
The built-in Demo mode is available to demonstrate operation of the device.
Disconnect the measuring cable from the device and turn the device off.
8.1.
To start Demo mode, press and hold the On/Off key
8.2.
second beep sounds and "Demo" indication appears on the screen.
During Demo cycle the operating sequence of the device is shown on the screen.
8.3.
Press On/Off key
When Demo cycle is completed, it is repeated automatically until interrupted by the
8.4.
operator.
To exit Demo mode press and hold the On/Off key
8.5.
sounds.
Note: If measuring cable is inserted into the device receptacle during Demo cycle, S5
Apex Locator exits Demo mode and switches automatically to regular operation mode.
9. Maintenance, Cleaning and Sterilization
9.1. General
• The device does not contain user serviceable parts. The service and repair should be
provided by factory trained service personnel only.
• All objects that were in contact with potentially infectious agents should be cleaned after
each use:
Lip clip, file clip and touch probe should be disinfected and sterilized by autoclaving
before the first use and between treatments. Please follow "Disinfection and sterilization
procedure" described in next section.
Measuring
cable,
cloth impregnated with aldehyde free disinfecting and detergent solution (a bactericidal
and fungicidal).
The measuring cable should not be autoclaved.
Use of agents other than specified above may cause damage to the
equipment and its accessories.
to pause the simulation; press On/Off key
the device and its cradle should be cleaned using tissue or soft
13
for about 1 sec. until the
again to resume.
for about 1 sec. until a beep