Ionizer (AX-ION-25)
The AX-ION-25 ionizer comes standard with the BA-6E / BA-6DE. The ionizer removes static electricity
by irradiating the target object with positive or negative ions, which are generated from four discharge
electrodes by DC corona discharge. Ordinarily, insulators such as powders, filters, and weighing paper
tend to be charged when the humidity is 45% RH or less, and an error of a few milligrams may occur
during weighing. By removing static from the weighing object with the ionizer, it is possible to eliminate
the error in the weighing value due to charging, and correct weighing can be performed. For details, refer
to the separate instruction manual in "1.1. Detailed manuals".
8.1. Using the ionizer
Step 1. Make sure that the AC adapter of the balance is not connected, and then connect the balance
and the ionizer (as noted in "2. Assembly and Installation").
Step 2. Connect the AC adapter, plug it in, and then turn on the power. The power lamp of the ionizer
lights up.
Step 3. Place the target object within the effective range of static elimination.
Step 4. Hold your hand over the IR sensor (touchless infrared proximity sensor) on the front of the
ionizer to start static elimination. The ION lamp indicating that static elimination is in progress
lights up.
Step 5. Refer to the figure below for the effective range of static elimination, and perform static
elimination. At factory setting, when 3 seconds have passed from the start of static elimination,
the ION lamp indicating that static elimination is in progress turns off and static elimination stops.
IR SENSOR (touchless infrared proximity sensor)
Power lamp (lights up when the power is turned on)
ION lamp (lights up during static elimination)
Effective range of static elimination
Top view
Side view
Effective range
0 cm
5 cm
Distance
Effective range
1 sec
Static elimination time
EN
7 cm
10 cm
1.5 sec
3 sec