Sensitivity Adjustment

6.1. Automatic sensitivity adjustment
This function automatically adjusts the sensitivity of the balance according to ambient temperature
change, set time or interval time using the internal weight. For details, refer to the separate instruction
manual listed in "1.1. Detailed manuals".
6.2. Sensitivity adjustment using the internal weight
Sensitivity adjustment using the internal weight can be performed with one key press.
Step 1. Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least
four hours for BA-6E/BA-6DE, with the AC adapter connected to the power supply.
Step 2. Press the CAL key. The balance displays
Step 3. The balance automatically performs sensitivity adjustment using the internal weight. Do not
apply vibration and the like to the balance.
Step 4. After sensitivity adjustment, the balance will output the "sensitivity adjustment report" and store
the data as set in the function table.
For details, refer to the separate instruction manual listed in "1.1. Detailed manuals".
Step 5. The balance returns automatically to weighing mode.
6.3. Sensitivity adjustment using an external weight
This function performs sensitivity adjustment of the balance using your external weight.
For details, refer to the separate instruction manual listed in "1.1. Detailed manuals".