4. Design and function
Function buttons
The gas density indicator is controlled via 4 function buttons, with each button having
a main and a secondary function. In general, the information printed on the buttons
corresponds to the main function: "ZERO", "LIGHT", "MENU", "ON/OFF". Once the
[MENU] button is active, the secondary function applies. These are, from left to right:
Cursor up "UP / ▲", cursor down "DOWN / ▼", left/right cursor "L / ◀" or "R / ▶" and
"HOME".
Pos.
Button
On/Off button
6
The main function is switching the gas density indicator on and off.
If the gas density indicator is already in menu mode, a short press on the [On/
Off] button brings up "HOME". A long press (at least 3 seconds) switches the
GDI-100-D off.
MENU button
7
Calling up the menu
By activating the [MENU] button, the menu mode is enabled. If the gas
density indicator is already in menu mode, depending on the display, "L" or
"R" will be executed. If the button is pressed for longer, after 2 seconds the
cursor switches (right ◀ or ▶ left).
Inputs are confirmed with the [MENU] button.
LIGHT button
8
Turning the backlighting on and off
By activating the [LIGHT] button (short or long press) the light is switched on.
The duration of the light being on depends on "LIGHT-OFF" in "SETTINGS".
■
■
If the GDI-100-D is in menu mode, with a short press on the [LIGHT] button,
the cursor can be moved downwards.
ZERO button
9
The current pressure value will be set to "0" (gauge) or reference pressure
(abs.).
By activating the [ZERO] button, the current pressure value is set to "0". A
maximum of 5 % of the measuring span can be corrected.
If the GDI-100-D is in menu mode, with a short press on the [ZERO] button,
the cursor can be moved upwards.
WIKA operating instructions, model GDI-100-D
1 x pressing the [LIGHT] button (Light = On)
2 x pressing the [LIGHT] button (Light = Off)
EN