Alternating mode: When ALTERNATE mode is selected, it enables the three-one
alternate function. Use the +/- button on the front panel to adjust comfort setting to
achieve the optimum patient comfort.
Time setting: Under alternating mode, the cycle time can be adjusted to 5, 10, 15, or 20
minutes per cycle by pressing time selection button on the front panel.
Static mode: When needed, the caregiver can select the STATIC mode for special
application. Under static mode, all cells are inflated with the same pressure setting and
the alternating effect stops. Use the +/- button on the front panel to adjust comfort
setting to achieve the optimum patient comfort.
Max function: Upon initial startup, use max function to allow mattress be inflated to
the maximum pressure. It will automatically return to previous setting 30 minutes after
activation, if it has not been manually changed to other modes.
Seat function: When seat function is turned on, it automatically increases pressure level
to support patient when needed, especially during head raised position and during
ingress/egress movements.
Lockout function: The pump is also equipped with a lockout function to prevent
unintentionally changing the setting. Caregiver can manually activate lockout function
by press and hold LOCK button for 3 seconds. To unlock, press and hold LOCK button
on the front panel for 3 seconds.
Power failure alarm: At any time when power failure occurs, both visual (amber color
LED will illuminate) and audio alarm will trigger.
NOTE: The low-pressure alarm function will not be activated until 45 minutes
after initial startup. This is to prevent faults alarm during initial inflation.
Mute function: The alarm can be temporarily muted by pressing the mute alarm button
on the front panel: it will mute the alarm for 20 minutes.
Maximum load / patient weight: The eMOTION Alternating Pressure Mattress System has a
maximum weight capacity of 200kg (440lb).
Low pressure alarm: For added safety, 5 minutes after low pressure is
detected, both visual and audio alarm will trigger.