The eMOTION is intended for use in the electromagnetic environment (for home healthcare and professional
healthcare) specified below.
The customer or the user of the eMOTION should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD) IEC
61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge IEC
61000-4-5
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
Power frequency
(50, 60 Hz)
magnetic field IEC
61000-4-8
NOTE
UT is the a.c. mains voltage prior to application of the test level.
* During DIP interference, the pump will outage these are normal. The pump outage does not affect the motor
operation.
Manufacturer's declaration-electromagnetic immunity
IEC 60601
Compliance level
test level
Contact:±8 kV
Contact:±8 kV
Air±2 kV,±4
Air±2 kV,±4
kV,±8 kV,±15 kV
kV,±8 kV,±15 kV
+ 2kV for power
+ 2kV for power
supply lines
supply lines
+ 1kV for
Not applicable
input/output lines
+ 0.5kV, +1kV
+ 0.5kV, +1kV
line(s) to line(s)
line(s) to line(s)
Not applicable
+ 0.5kV, +1kV,+
2kV line(s) to
earth
Voltage dips:
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 0,5 cycle
0 % UT; 1 cycle
0 % UT; 1 cycle
70 % UT; 25/30
70 % UT; 25/30
cycles
cycles
Voltage
Voltage
interruptions:
interruptions:
0 % UT; 250/300
0 % UT; 250/300
cycle
cycle
30 A/m
30 A/m
50 Hz or 60 Hz
50 Hz, 60 Hz
Electromagnetic environment-guidance
(for home healthcare and professional
healthcare environment)
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment. If the user of the eMOTION requires
continued operation during power mains
interruptions, it is recommended that the
eMOTION be powered from an uninterruptible
power supply or a battery.
The eMOTION power frequency magnetic fields
should be at levels characteristic of a typical
location in a typical home healthcare and
professional healthcare environment.