Resonance R25C Manuel D'utilisation page 57

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

AUTORUN
It appears when middle ear
CHECK
analyzer tests are performed
when the Autorun function has
been activated
DO YOU
This query appears after entry of
CONFIRM ?
plot cancellation or parameter
change requests
CHANGE
This message appears at the end
CALIBRATION
of the device calibration proce-
DATE ?
dure, once the "SAVE" softkey has
been pressed and values have
been changed.
INSERT PROBE
This message appears when
INTO THE
on the Compliance Calibration
SPECIFIED CAVITY
page
PRESS
This message comes up on the
CONFIRM/AUTO-
Compliance Calibration page
MATIC
the instant in which the probe is
KEY
inserted into the specified cavity
CHECK CAVITY
This message is given on the
Compliance Calibration page if
the probe is not inserted into the
displayed cavity
READ
This message is given on the
VALUE (ml)
Compliance Calibration page, at
the end of the calibration process
INSERT PROBE
This message appears on the
INTO THE 2cc
ARLT Calibration page, if the
CAVITY
probe is not inserted into the
2cc cavity.
PRESS RUN TO
This message appears on the
PERFORM
ARLT Calibration page, when
CALIBRATION
the probe is inserted into the
2cc cavity
ENTER
This message appears when
PASSWORD
attempting first-time access to
the device calibration procedure,
once the unit has been switched
on
WRONG
This message appears when
PASSWORD
attempting to access the device
calibration procedure using a
wrong password
System
It means that the test is
message
being performed with
the Autorun function
activated
System
It means that a plot
message
cancellation or a
parameter change
request has been
entered
System
It provides the option
message
of updating periodic
calibration dates
System
This message refers
message
to the cavity value
displayed on the Cali-
bration page, that it is
necessary to insert the
probe into for calibra-
tion performance
System
It provides instruc-
message
tions that need to be
followed in order to
proceed with compli-
ance calibration
System
It means that the probe
message
has been inserted into
the wrong cavity
System
It means that the value
message
that is displayed is the
real-time value reading
by the device
System
You are instructed to
message
insert the probe into
the 2cc cavity in order
to perform calibration
System
It provides instructions
message
on proceeding with
calibration
System
It instructs you to enter
message
your access password
to unlock the device
calibration procedure
System
It means that you have
message
entered the wrong
password for access to
the Calibration menus
No action is required–if
the Autorun function
is not wanted, it can be
deactivated via the Setup
menu
Press either the YES or
the NO when the query
appears
Press either the YES or
the NO when the query
appears
Insert the probe into the
test cavity as displayed
Press the confirmation
key to calibrate selected
probe tone frequency
or automatic key to
calibrate all probe tone
frequencies (PT models)
Proceed with inserting
the probe into the
correct test cavity, as
displayed
No action is required
Proceed with inserting
the probe into the 2cc
cavity
Press the Start key as
instructed
Enter your password via
the softkeys
Press the Enter key, then
re-enter the password
correctly
57