TECHNICAL SPECIFICATION
• Measuring range:
Human Body: 34°C ~43°C (93.2°F ~ 109.4°F)
• Laboratory Accuracy:
34°C ~43°C±0.2°C (93.2°F ~ 109.4°F ±0.4°F)
• Measuring site: Ear
• Reference body site: Oral
• Calibration Accuracy:
±0.2°C(±0.4°F): from 35°C ~42°C (95°F ~ 107.6°F)
±0.3°C(±0.5°F): out of range
• Display resolution: 0.1°C/°F
• Operating environment:
15°C~35°C (59°F ~ 95°F) with relative humidity up to 95% (non condensing)
• Storage/transportation environment:
-25 to 55 °C(-13 to 131 °F) with relative humidity up to 95% (non condensing)
• Power supply: A lithium 3V CR2032 battery
• Memory: 10 sets
• Weight: approx. 49g (with battery)
• Dimensions: approx. 112.8mm×34.4mm×44.5mm (L×W×H)
• Shelf life: 3 years
EMC TABLES
Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information.
1. Electromagnetic/Radio Frequency Interference (EMC/RFI): Readings may be
affected if the unit is operated within a radio frequency electromagnetic
field strength of approximately 3 volts per metre, but the performance of the
instrument will not be permanently affected. Care should be taken to keep the
thermometer at least 15cm/6 inches away from R/C transmitters to avoid radio
frequency interference.
2. Avoid keeping the thermometer too close to objects that continuously generate
high heat (like a hot plate) for long periods of time, which can cause overheating
of the thermometer.
10 | GB