Modbus
Starting
Size
Modbus
address
[bits]
Access
<n>220
16
holding
register
<n>230
32
holding
register
<n>240
32
holding
register
<n>250
32
holding
register
<n>260
32
holding
register
<n>270
32
input register integer Power / AD value
<n>280
32
holding
register
<n>290
32
holding
register
<n>300
16
holding
register
<n>310
32
holding
register
<n>320
32
holding
register
104
Data
Content
Type
char
Emissivity Source
float
Presel. Emissivity
float
Valley hold time
float
Average time
float
Peak hold time
float
Presel. Setpoint
float
Transmissivity
short
FTC3 trigger/hold
float
Setpoint relay function
float
Adv. hold hysteresis
Rev. G2 Jul 2017
Values
I: internal command
E: ext. input (0V .. 5V)
D: digital selected FTC1-3
0.1 .. 1.1
0.0 .. 998.9s (999: infinite)
0.0 .. 999.0 [seconds]
0.0 .. 998.9 [seconds] (999: infinite)
0.1 .. 1.0
1: trigger, 2: hold
number within range <n>060 .. <n>070 XS
MI3[M]
command
ES
EV
F
G
P
Q
SV
XG
XN
XY
MI3