Apex DOMUS AUTO Mode D'emploi page 9

Matelas pour traiter et prévenir les escarres
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
English
3. THERAPY
The Therapy button allows you to choose one of the therapy modes. There are four
selectable therapy modes available. Continually press Therapy button to select a desired
mode:
MaxFirm
Alternate
A.
MaxFirm
The surface will go into Max firm automatically when the power switch is turned on. This
insures the pump is able to reach its maximum operating pressure. Once the max
pressure level is reached, the pump will automatically switch into alternating mode. User
can also use this function during patient transfers or nursing procedures for better
support. This mode will last for 20 minutes and then it will return to previous setting
automatically. To disable this function, press Maxfirm button again.
B.
Alternate
The mattress will alternate every 10 minutes under this therapy mode.
C.
Static
Non-alternating mode, all of the air cells is equally inflated.
D.
Seat Inflate
The seat inflate features additional supports to the patient during upright position.
4. Alarm Mute
Press alarm mute button to temporary suspend the low-pressure alarm (LED light and
Should the situation not resolved within 5 mins, the alarm shall resume to notify
A.
PFA (Power Failure Alarm)
During power failure situation, the Power Failure LED light will lit on with buzzer. By
pressing the mute button to disable both buzzer and LED.
B.
Low Pressure Indicator
When low pressure LED lights up, the pressure inside of air mattress is below normal.
Please refer to troubleshooting.
C.
Service
This feature will light during mechanical failure situation. User can notify the technician
for repair.
Static
Alternate+Seat Inflate
7
Static+Seat Inflate
buzzer).
the patient.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

9p-077520

Table des Matières