Télécharger Imprimer la page

Masimo RD rainbow SET MD20 Série Mode D'emploi page 3

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 6
RD rainbow SET™ MD20 Series
Patient Cables
Reusable
INDICATIONS
The RD rainbow SET™ MD20 Series patient cables have the same indications for use as the applicable sensors. Please see
the compatible sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
The RD rainbow SET MD20 Series patient cables are used with RD rainbow SET Series sensors.
RD rainbow SET MD20 Series patient cables are for use only with devices containing Masimo rainbow SET technology
Version 7.1 or higher or licensed to use rainbow compatible sensors. Consult individual oximetry system manufacturers
for compatibility of particular devices and sensor models. Each device manufacturer is responsible for determining
whether their devices are compatible with each sensor model.
RD rainbow SET MD20 Series patient cables have been validated using Masimo rainbow SET® technology.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed
to use Masimo sensors.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable
and discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable
and sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable may result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or
disconnecting either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize
the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may
damage the electrical components, potentially leading to patient harm.
• Caution: Replace the patient cable when a replace cable or a persistent low SIQ message is displayed when the
patient monitoring time is exhausted.
• Note: The patient cable is provided with X-Cal™ technology to minimize the risk of inaccurate readings and
unanticipated loss of patient monitoring. The cable will provide 17,520 hours of patient monitoring time. Replace
the cable when the patient monitoring time is exhausted.
INSTRUCTIONS
A) Attach the RD rainbow SET MD20 Series patient cable to the Device
1. Orient the connector of the cable to mate with the patient cable connector on the Module.
B) Attaching the RD rainbow SET MD20 Series patient cable connector to an RD SET disposable sensor connector
1. Refer to Fig. 1. Orient the sensor connector to the patient cable connector as shown.
2. Refer to Fig. 2. Insert the sensor connector completely into the patient cable connector until it locks in place
as shown.
C) Disconnecting the RD rainbow SET MD20 Series patient cable connector from the RD SET disposable sensor
connector
1. Refer to Fig 3. Pull firmly on the sensor connector to remove it from the patient cable.
DIRECTIONS FOR USE
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
3
en
Non-Sterile
9000C-eIFU-1115

Publicité

loading