DISINFECTION AND STERILIZATION
Disinfection of and sterilization of should be carried out only after procedure cleaning of the product.
For disinfection products it is recommended to use disinfectant solution, for example, Gigasept FF NEU or
Bomix Plus. When choosing a disinfectant means consider List of disinfectants VAH (www.vah-online.de).
Please follow the instructions provided by the manufacturer of the selected agent. Acceptable use 70%
solution ethanol method wipe.
Do not soak the product in a disinfectant solution.
Do not use disinfectants with means of mass content shares of isopropyl alcohol more than 5%.
Gas sterilization with ethylene oxide is acceptable.
Do not use autoclave or other hot sterilization methods.
STORAGE
Temperature: from +5°С to +40°С. Humidity: from 20% to 80%.
Don`t expose direct sunlight.