1. Photometer PF-12
1.1 Technical Data
Type:
Wavelength range:
Optics:
Wavelengths:
Wavelength accuracy:
Light source:
Detector:
Blank value:
Measuring modes:
Photometric range:
Photometric accuracy:
Stability:
Cuvette holder:
Data memory:
Display:
Operation:
Quality control:
Interfaces:
Update:
Operating range:
Power supply:
Housing:
Dimensions:
Weight:
Warranty:
This device complies with the following directives:
- 2006/95/EC - Low-Voltage Directive
- 2004/108/EC - EMC Directive
Single-beam filter photometer with microprocessor
control, self-test and auto-calibration
340 – 860 nm
A utomatic filter wheel with 7 interference filters, Insen-
sitive to external light for fast measurements without
cuvette slot cover
345 / 436 / 470 / 540 / 585 / 620 / 690 nm
plus 1 compartment for an additional filter
± 2 nm
Bandwidth at half of maximun transmission 10 – 12 nm
Tungsten lamp
Silicon-photodiode
Automatic
Over 100 preprogrammed tests (NANOCOLOR
tests und VISOCOLOR
transmission, factor, standard,
10 freely programmable methods
± 3 E
± 1%
< 0,002 E/h
Round tubes 16 mm OD
200 results, GLP-compatible
Backlit graphic display, 64 x 128 pixels, 12 languages;
All important data at a glance: result with measuring unit,
date, time, sample number, sample location, dilution
Intuitive menu navigation, durable keypad, test selec-
tion via entering the test number or from parameter lists
With NANOCONTROL NANOCHECK
USB 2.0
Free of charge via Internet/PC
0 – 50 °C, up to 90 % relative humidity
USB-power supply, disposable or rechargeable batteries
Waterproof, IP 67
215 x 100 x 65 mm
0.7 kg
2 years
ECO tests), absorbance,
®
tube
®
13