± 2 kV for power
Electrical fast
cord
transient burst
± 1 kV for input/
output lines
± 1 kV differential
mode voltage
Surge
± 2 kV common-
mode voltage
<5% UT for 0.5
cycle (> 95% dips
Voltage
on UT)
dips, short
40% UT for 5
interruptions
cycles (60% dips
and voltage
on UT)
changes in
70% UT for 25
power input
cycles (30% dips
line
on UT)
<5% UT for 5s (>
95% dips on UT)
Power
frequency
3 A/m
magnetic field
(50/60Hz)
Note: UT refers to the AC voltage before applying the test voltage.
Not applicable Not applicable
Not applicable Not applicable
Not applicable Not applicable
The power
frequency magnetic
field should have
the horizontal
characteristics of
3 A/m, 50/60
power frequency
Hz
magnetic field of
a typical place in a
typical commercial
or hospital
environment.
15
(3)
Guide and manufacturer's statement - Electromagnetic immunity
This device is intended for use in the following electromagnetic
environment. The purchaser or user of this device should ensure that it
is used in this electromagnetic environment:
Immunity
IEC60601 test
Compliance
test
level
level
3 V (effective
RF
value)
Not
conduction
150 kHz ~ 80
applicable
MHz
3 V/m 80 MHz
RF radiation
3 V/m
~ 2.5 GHz
Electromagnetic
environment - Guide
Portable and mobile
RF communication
equipment should
not be used near any
part, including cables,
of the product at a
distance shorter than the
recommended isolation
distance. This distance
should be calculated with
the formula corresponding
to the transmitter
frequency.
Recommended isolation
distance
d=1.2 P
80 MHz~800 MHz
d=2.3 P
800 MHz~2.5 GHz
16