- Do not sterilize the device in an autoclave, with ethylene oxide or by immersing it in liquid. The
device must not be sterilized.
7. TECHNICAL SPECIFICATIONS
Product name
BewellConnect® MyOxy
Model
BW-OX1
Classification
Classe IIa
Applied part type BF
IP22
Display type
LED screen
LED Specifications
Wavelength: Red = 660 ±3nm - Infrared = 905 ±10nm
Radiant power : Red = 3.2mW - Infrared = 2.4mW
Dimensions
32 x 32 x 62 mm / 1.3 x 1.3 x 2.4 in
Weight
40 g / 1.41 oz
Display and
SpO
: Display range : 35 - 99% / Measuring range : 70 - 99%
2
measuring ranges
Pulse: Display and measuring ranges: 30 - 250 bpm
Accuracy
SpO
: 70 - 100% : ± 3%
2
Pulse : ± 2 bpm
Resolution
SpO
: 1%
2
Pulse : 1 bpm
Normal working
Temperature : 5°C - 40°C / 41°F - 104°F
conditions
Humidity : 15-93%RH
Storage conditions
Temperature : -25°C - 70°C / -13°F - 158°F
Humidity : ≤93%RH
Atmospheric pressure : 86-106Kpa
Technology
Bluetooth 4.0 Smart Ready
Battery
Lithium 280mAh
Autonomy
400 readings
Automatic shut-
8 seconds
down
22
BewellConnect - BW-OX1 - User Manual - 072017v2
Bluetooth
10m
transmission
distance
Frequency
2400 - 2483.5MHz
USB
USB port 2.0
Accessories
Pulse oximeter, instruction manual, protective pouch, USB cable
provided
8. TROUBLESHOOTING
If one of the following issues arises when using your MyOxy, consult this trouble-shooting
guide. If the problem continues, please contact our customer service department at +33 892
350 366 (0,40€/mn).
ISSUE
POSSIBLE CAUSE
Sp0
or PRbpm
Your finger must be
2
do not appear
correctly inserted.
correctly
SpO
reading
2
too small to be
measured.
Sp0
or PRbpm
Your finger may not
2
are unstable
be inserted deeply
enough.
The user's finger is
trembling or the
user is moving.
Impossible to
Battery low
turn on the
Oximeter is damaged Contact customer service
device
BewellConnect - BW-OX1 - User Manual - 072017v2
SOLUTION
Change the position of your finger and try again.
Try several times. If you have any doubts about
a problem related to the product, seek medical
attention
to obtain an accurate diagnosis.
Change the position of your finger and try again.
Try not to move.
Recharge the device with the USB cable
23