4.6.2 Password Protected Menus
The customer settings sub-menus contain the parameters that may be customised by the end user. To prevent unauthorised changes being
made these menus are password protected and cannot be accessed until the password has been entered correctly.
Use the
key to navigate to page 3 "Customer Settings".
4.6.3 Hour Meters
4.6.4 Fault Log
The following codes are used within the system:
Fault Codes
C1
P1
P2
E1
Y1
Y2
Y3
Y4
Y5
S1
Note. Any faults that are active when the power is switched off, and remain active when the power is re-applied, will cause a new entry to be
added into the fault log.
16
To enter the password from the main operating menu press and hold both the
seconds until the menu changes to the password prompt as shown
The flashing cursor will be positioned over the first digit.
Using the
key change the first digit of the code and press
Repeat the process and enter the following password 1 2 1 _ _.
When the password has been entered correctly the Hour Meters menu will be displayed.
The time in hours that the generator has been producing gas.
The time in operating hours that the generator can produce gas before a service is required.
The time in hours that the generator has been operating in Economy mode.
The fault log contains details of the ten most recent faults that have occurred on the generator. Each fault is
represented by a code which is displayed along with the time (hours run) at which the fault occurred and the
status of the fault.
Pressure Start Inhibit
Inlet Pressure Fault
Pressure Sensor Fault
Power Failure
O
Alarm
2
O
Communication failure
2
Incorrect cell selected
O
high (out of range)
2
O
zero drift error
2
Service due
Notes
Low inlet pressure. Inhibits start.
Low inlet pressure during cycling.
Pressure sensor communication error.
Communication fault between O
2
Occurs when O
> 25% (% generators) / O
2
Contact Parker domnick hunter
and
keys for approximately 5
. The cursor will move to the next digit.
analyser and control board
> 1.05% (ppm generators)
2