MultiCap user manual
Electrodes and impedances
1. In order to minimize the impedance, the
electrode position on the scalp should be
prepared. We recommend the abrasive gel
"EVERY" or "NUPREP" which both have very
good
qualities
user-friendly basecut of the MultiCap, all
electrode positions can be reached quite simply.
2. Put the electrodes onto the desired positions.
For this, press down the electrode into the
electrode- body until it´s be fixed.
3. Make sure, that the cables of the electrodes are
not
tangled
or
velcro-strap to focus the cables (see photo).
4. In order to raise the contact-surface, use a
conductive gel like "NEURGEL", or for long time
recordings the "SYNAPSE" gel that you can fill
with the provided blunt needle kit into the
intended gap (notch) of the electrode.
Electrodes and impedances
1. In order to minimize the impedance, the
electrode position on the scalp should be
prepared. We recommend the abrasive gel
"EVERY" or "NUPREP" which both have very
good
qualities
user-friendly basecut of the MultiCap, all
electrode positions can be reached quite simply.
(MultiCap-Base)
for
preparation.
folded.
Use
the
(MultiCap-Soft)
for
preparation.
By
the
provided
By
the
CAP
CAP
15