MultiCap user manual
Procedure
1. Unpack the product and check the completeness and appearance of all parts.
In case of missing or damaged parts please contact your supplier.
2. Fill the syringe with gel and install the supplied blunt needle. The needle
should sit tightly on the syringe.
Use only the supplied blunt needle of plastic. A sharp needle of matallic can
Instruction for use
1. Measure the head circumfence of your patient and choose the proper cap
size.
2. Pull the cap on the head of the patient. Beginning from the forehead, where
the subject may be fixed, pull the cap over the head. Next, adjust the electrode
on position Cz at the half way between Nasion and Inion. The frontpolar and
occipital electrode should now be placed correctly. Otherwise choose another
cap size. When measured correctly, the frontpolar electrodes (Fp1/Fp2) should
lie directly above the eyebrows.
3. Make sure, that the position Cz is centered between the left and the right
side, as well that the lateral electrode position is symmetrical.
4. Fix the cap with the chin strap and make sure that the cap fits right. The wide
part of the belt should fasten at the chin, not at the neck. The belt tension should
provide a reliable fixing of cap.
Don`t pull the belt too tight. The patient should feel
14
Attention!
damage the skin of the patient.
Attention!
comfortable even the recording
CAP
CAP