Accu-Chek Ultraflex Mode D'emploi page 8

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Avoid mechanical stress on the infusion site
and on the Accu-Chek Ultraflex. Pay attention
when carrying heavy weights.
Prepare the infusion site before insertion ac-
cording to CDC (Centers for Disease Control
and Prevention) guidelines or institutional
protocol.
Check blood glucose level 1–3 hours after
inserting your Accu-Chek Ultraflex and inspect
the infusion site on a regular basis. Do not
insert a new soft cannula just prior to bedtime.
Check blood glucose frequently (check with
your physician).
Check your infusion set at the luer-lock to
tube connection periodically (every  hours

during the day and before going to sleep)
during use. A break may result in leakage,
causing interruption of insulin delivery. If
you notice a break, change your infusion
set immediately and check your blood glu-
cose.
lf blood glucose level gets unexplainably too
high, or an «occlusion» alarm occurs, check
for clogs and leaks. lf in doubt, change your
Accu-Chek Ultraflex.