WARNING
Ensure clean application. Do not touch the soft
cannula or introducer needle. Make sure that the
soft cannula is fully inserted into the subcutane-
ous tissue and not kinked at the skin surface.
Once the introducer needle punctures the skin,
the soft cannula must be quickly pushed in the
rest of the way, or it may kink upon insertion,
causing reduced or blocked insulin delivery.
Picture 5
Carefully smooth out the self-adhesive (F) onto
the skin and run your fingers over the self-adhe-
sive to ensure good skin contact is achieved.

Picture 6
Hold the self-adhesive (F) in place, gently press
the side clips of the blue holder cap (M) and pull
it from the cannula housing (G).
Picture 7
To remove the introducer needle (C) hold the
cannula housing (G) firmly to your body. Place
one finger on the top of the cannula housing
and withdraw the introducer needle in a straight
manner using the introducer needle handle (A).
Remember to dispose the introducer needle (C)
safely (needle container). You may use the pro-
tective cover (B) provided to cover the needle.
Use caution when covering the steel needle;