Product specifications

Displayed temperature range:
Operating ambient temperature range:
Display resolution:
Accuracy for displayed temperature range: ±0.4 °F (95.9–107.6 °F) (35.5 °C–42 °C)
Clinical repeatability:
Long term storage ranges
Temperature:
Humidity:
Shock:
Measurement time:
Automatic «power down»:
Battery life:
Battery type:
Thermometer dimension:
Thermometer weight:
This infrared thermometer meets requirements established in ASTM Standard E 1965-98
(for the thermometer system [thermometer with probe cover]). Full responsibility for the
conformance of the product to the standard is assumed by Kaz USA, Inc.,
250 Turnpike Road, Southborough, MA 01772, USA.
ASTM laboratory accuracy requirements in the display range of 96.8 °F to 102.2 °F (36 °C to 39 °C) for
infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas for mercury-in-glass and electronic thermometers,
the requirement per ASTM Standards E 667-86 and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).
A clinical summary is available upon request.
This device conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers» –
This products conforms to the provisions of the EC directive 93/42/EEC
(Medical Device Directive).
To protect the environment, dispose of empty batteries according to national or
local regulations.
14
68–108 °F (20–42.2 °C)
50–104 °F (10–40 °C)
0.1 °F or °C
±0.5 °F (outside this temperature range)
± 0.26 °F (± 0.14 °C)
–4 °F to 122 °F (–20 °C to 50 °C)
95% non condensing
withstands drop of 3 feet
Initial start-up time: 3–4 seconds
Measurement time: 2–3 seconds
60 seconds
6 months/1000 measurements
2 x MN 1500 or 1.5 V AA (LR6)
6˝ x 1.7˝ x 1.3˝ (152 mm x 44 mm x 33 mm)
3.6 oz (100 g) without batteries
TYPE BF
EQUIPMENT
Internally powered equipment.
Continuous operation.
Not protected against
ingress of water.
Service
The Braun ThermoScan thermometer has a limited three year warranty. Should service be required during
or after the warranty period you must contact Welch Allyn at one of the locations listed on the inside cover
to obtain a return authorization number and shipping address. Repackage the thermometer carefully in
its original packaging or securely packed to avoid damage during shipping. Include the original sales slip
indicating the date of purchase, a note describing the problem, your authorization number, and your return
address. Send the Braun ThermoScan thermometer prepaid and insured.
Note: Shipping address varies by country.
Answers to any questions regarding operation of the Braun ThermoScan thermometer may be obtained by
contacting Welch Allyn at one of the numbers listed on the inside cover.
Limited three year manufacturer's warranty
For Model PRO 4000
The PRO 4000 instrument is warranted by Kaz to be free from defects in material and workmanship under
normal use and service for a period of three years from the date of delivery to the first professional user who
purchases the instrument.
This warranty does not cover batteries, damage to the probe window, or damage to the instrument caused
by misuse, negligence or accident, and extends to only to the first purchaser of the product. Additionally
this warranty becomes void, if the thermometer is operated with anything other than genuine Braun
ThermoScan
Probe Covers.
®
Thermometer service or replacement will be carried out by Kaz subject to the terms of this warranty.
Loss or damage in return shipment to Kaz shall be at purchaser's risk. Kaz reserves the right to replace a
non-conforming product with the most comparable product currently available. This product has a
five year minimum expected lifetime.
IN NO EVENT SHALL KAZ BE LIABLE FOR ANY SPECIAL, INCIDENTAL, INDIRECT OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE PURCHASE OR USE OF ANY PRODUCT. This warranty shall not
apply to, and Kaz shall not be responsible for, any loss arising in connection with the purchase or use of any
product which has been serviced by anyone other than a Kaz authorized service representative or altered in
any way so as, in Kaz's judgment, to affect its stability or reliability, or which has been subject to misuse or
negligence or accident, or which has the serial number or lot number altered, effaced or removed, or which
has been used other than in accordance with the instructions furnished by Kaz. IN NO EVENT SHALL KAZ'S
LIABILITY EXCEED THE ORIGINAL COST OF THE PRODUCT.
This warranty is in lieu of all other warranties, express or implied, and of all other obligations or liabilities on
Kaz's part, and Kaz neither assumes nor authorizes any representatives or other person to assume for it any
0297
other liability in connection with the sale of its products.
TO THE EXTENT PERMITTED BY LAW, KAZ DISCLAIMS ALL STATUTORY AND IMPLIED WARRANTIES,
INCLUDING BUT NOT LIMITED TO, WARRANTIES OF MERCHANTIBILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
W E L C H A L LY N A U S T R A L I A P T Y LT D
U N I T 5 3 8 S O U T H S T R E E T
R Y D A L M E R E , N S W 2 116
A U S T R A L I A
10HK
Medical Equipment with
respect to electrical shock,
fire and mechanical hazards
only in accordance with
UL60601-1
15